Brand
SonarMed Inc
4 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for SonarMed Inc and translate it into plain English — free weekly digest, optional same-day alerts.
- 2024-04-24FDA-DeviceClass IAW-M0001 SONARMED MONITOR. Used in conjunction with SonarMed Acoustic Sensors for airway monitoring.
Software anomaly resulted in failure to detect a partial obstruction in 2.5 mm sensors and up to 3mm distal to the sensor.
- 2024-04-24FDA-DeviceClass IAW-S035 NEO SONARMED SENSOR 3.5MM. Used in conjunction with SonarMed Monitor for airway monitoring.
The failure to detect the partial obstruction in a 2.5mm sensor.
- 2024-04-24FDA-DeviceClass IAW-S030 NEO SONARMED SENSOR 3.0MM. Used in conjunction with SonarMed Monitor for airway monitoring.
The failure to detect the partial obstruction in a 2.5mm sensor.
- 2024-04-24FDA-DeviceClass IAW-S025 NEO SONARMED SENSOR 2.5MM. Used in conjunction with SonarMed Monitor for airway monitoring.
The failure to detect the partial obstruction in a 2.5mm sensor.
Get notified about new SonarMed Inc recalls
Free weekly digest. We email when a new federal recall affects SonarMed Inc or anything else in your household.