- 2025-04-02FDA-DeviceClass IICardioMEMS Heart Failure (HF) System that consist of: -CardioMEMS Hospital System Model CM3100 -CardioMEMS Backend Web Application Model CM4000 -Merlin.net Patient Care Network (PCN) Heart Failure Web Application Model MN6000 The CardioMEMS" HF System provides pulmonary artery (PA) hemodynamic data used for monitoring and management of heart failure (HF) patients. The system measures changes in pulmonary artery (PA) pressure which physicians use to initiate or modify heart failure treatmen
Due to issues associated with data migration of patient information and results to a Cloud service, the firm has received complaint associated with some patient profile information being duplicated while other patient information is missing.
- 2025-01-29FDA-DeviceClass IICARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Hosptial Electronics System (PES) REF CM3000 Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Which interfaces with CardioMEMSTM HF System Patient Electronics System (PES) and CardioMEMSTM HF System Hospital Electronics System (HES)). Model Number MN6000
As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
- 2025-01-29FDA-DeviceClass IICARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Patient Electronics System (PES) REF CM1100 Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Which interfaces with CardioMEMSTM HF System Patient Electronics System (PES) and CardioMEMSTM HF System Hospital Electronics System (HES)). Model Number MN6000
As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
- 2025-01-29FDA-DeviceClass IIMerlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with CardioMEMS" Hospital System model CM3100) The CardioMEMS" HF System provides pulmonary artery (PA) hemodynamic data used for monitoring and management of heart failure (HF) patients. The system measures changes in pulmonary artery (PA) pressure which physicians use to initiate or modify heart failure treatment.
Due to a data migration from on-premises server to the cloud environment, some customers reported that they were not able to transmit readings.
- 2024-07-24FDA-DeviceClass IISt. Jude Medical Agilis NxT Steerable Introducer, REF 408309
One lot of product has dilators that are too short and will not extend outside the introducer sheath.
- 2023-11-15FDA-DeviceClass IICardioMEMS Patient Electronic System (PES), Model Number CM1100. Used for wirelessly measuring and monitoring pulmonary artery pressure and heart rate.
Potential for damaged and frayed power connector plug with repeated bending or manipulation beyond 90 degrees.
- 2017-10-18FDA-DeviceClass IFortify VR - Model Numbers: CD1231-40 and CD1231-40Q; Fortify ST VR, Model Numbers: CD1241-40 and CD1241-40Q; Fortify Assura VR, Model Numbers: CD1257-40, CD1257-40Q, CD1357-40C, CD1357-40Q; Fortify Assura ST VR, Model Numbers: CD1263-40, CD1263-40Q, CD1363-40C, CD1363-40Q; Fortify DR, Model Numbers: CD2231-40 and CD2231-40Q; Fortify ST DR, Model Numbers: CD2241-40 and CD2241-40Q; Fortify Assura DR, Model Numbers: CD2257-40, CD2257-40Q, CD2357-40C, CD2357-40Q; Fortify Assura ST DR, Model Numbers
The device may exhibit premature battery depletion.