7 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Surgical Theater Inc and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-11-05FDA-DeviceClass IISurgical Planner, BrainSTORM, software version BST.9.3.0, orthopedic augmented reality
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
- 2025-11-05FDA-DeviceClass IISpineAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SPR.1.2.0, orthopedic augmented reality
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
- 2025-11-05FDA-DeviceClass IIEndo Surgical Navigation Advanced Platform Patient Engagement (EndoSNAP), software version SNP.1.12.0, imaging processing radiological system
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
- 2025-11-05FDA-DeviceClass IISyncAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SNP.1.12.0, imaging processing radiological system
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
- 2025-11-05FDA-DeviceClass IISurgical Navigation Advanced Platform Patient Engagement (PE SNAP), software version SNP.3.12.0, imaging processing radiological system
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
- 2025-11-05FDA-DeviceClass IISurgical Navigation Advanced Platform (SNAP), software version SNP.3.12.0, imaging processing radiological system
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
- 2025-11-05FDA-DeviceClass IISuRgical Planner (SRP), software version SRP.7.12.0, imaging processing radiological system
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.