Thoratec Corp.
6 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Thoratec Corp. and translate it into plain English — free weekly digest, optional same-day alerts.
- 2024-06-19FDA-DeviceClass IThoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN, 1286INT and L1286INT. Left ventricular assist device
System monitor screen may display atypical behavior: Overlapping screens/buttons, Frozen screen, Distorted text, blanks, or zeroes in place of values, Unresponsive buttons where the user is unable to initiate a command, which may lead to anxiety /inconvenience, hemodynamic compromise, extended surgical time, and Pump Stop command may be selected
- 2024-05-08FDA-DeviceClass IThoratec HeartMate 3, LVAS Implant Kit, REF:106524INT, Sterile EO, RxOnly
Potential for a leak path at the seal interface between the left ventricle assist system inflow cannula and the titanium apical cuff.
- 2024-05-08FDA-DeviceClass IThoratec HeartMate 3, LVAS Implant Kit, REF:106524US, Sterile EO, RxOnly
Potential for a leak path at the seal interface between the left ventricle assist system inflow cannula and the titanium apical cuff.
- 2024-03-27FDA-DeviceClass IHeartMate 3 Left Ventricular Assist System (LVAS): 1. HeartMate 3 Sealed Outflow Graft with Bend Relief, Investigational Device Exemption (IDE) 2. HeartMate 3 Sealed Outflow Graft with Bend Relief, US 3. HeartMate 3 Sealed Outflow Graft with Bend Relief, OUS 4. HeartMate 3 LVAS Implant Kit, IDE 5. HeartMate 3 LVAS Implant Kit, US 6. HeartMate 3 LVAS Implant Kit, OUS HeartMate 3 Left Ventricular Assist System: The HeartMate 3 LVAS is designed to provide therapeutic benefit to those affec
Observed outflow graft deformation known as Extrinsic Outflow Graft Obstruction (EOGO) associated with Left Ventricular Assist System.
- 2024-03-27FDA-DeviceClass IHeartMate II Left Ventricular Assist System (LVAS): 1. HeartMate II Sealed Outflow Graft with Bend Relief REF 103393 2. HeartMate II LVAS Implant Kit, US REF 106015 3. HeartMate II LVAS Implant Kit, OUS REF 106016 Update (February 14, 2024) HeartMate II Left Ventricular Assist System (LVAS): 1. HeartMate II LVAS Implant Kit, OUS REF 102139 2. HeartMate II LVAS Implant Kit, US REF 1355 3. HeartMate II LVAS Implant Kit (with RSOC Controller), US REF 103695 4. HeartMate II LVAS Implant K
Observed outflow graft deformation known as Extrinsic Outflow Graft Obstruction (EOGO) associated with Left Ventricular Assist System.
- 2024-02-14FDA-DeviceClass I1. HeartMate Touch Communication System, REF HMT1100 2. HeartMate Touch Communication System, Rental REF HMT1100-R 3. HeartMate Touch Communication System, HMT1150 The HeartMate Touch Application is loaded onto a tablet and will provide a user interface with state-of-the-art hardware/software platforms for supporting LVAD patients. Along with being able to program and monitor the LVAS, the HeartMate Touch Application introduces features that enables users to setup pairing and connection wit
Due to software and controller systems that results in unexpected pump start or pump stop.