Ventec Life Systems, Inc.
2 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Ventec Life Systems, Inc. and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-03-05FDA-DeviceClass IVOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002
Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.
- 2024-03-06FDA-DeviceClass IVOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.