- 2025-06-18FDA-DeviceClass IZ-800WF Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.
Unreleased software versions were installed on distributed devices without verification or validation.
- 2025-06-18FDA-DeviceClass IZ-800W Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.
Unreleased software versions were installed on distributed devices without verification or validation.
- 2025-06-18FDA-DeviceClass IZ-800F Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.
Unreleased software versions were installed on distributed devices without verification or validation.
- 2025-06-18FDA-DeviceClass IZ-800 Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.
Unreleased software versions were installed on distributed devices without verification or validation.
- 2024-12-04FDA-DeviceClass IIIBrand Name: Z-800 Infusion System Product Name: Z-800WF Model/Catalog Number: Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision physician or other certified healthcare professional.
For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.
- 2024-11-27FDA-DeviceClass IIZ-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision of a physician or other certified healthcare professional.
Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.
- 2024-10-16FDA-DeviceClass IZ-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W, Z-800WF; Software Version: Z-800 6.1.01 and 6-1.07z; Z-800F 4.1.02 and 4.1.08z; Z-800W 3.1.32 and 3.1.64z, Z-800WF 3.1.32 and 3.1.64z
There is a defect in the air-in-line software algorithm.
- 2024-07-10FDA-DeviceClass IIZyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pumps
Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unplugged from the AC mains power.
- 2024-03-06FDA-DeviceClass IIZyno Medical Z800 Large Volume Infusion Pumps -Intended to provide intravenous infusion of parenteral fluids, blood and blood products to a patient under the direction or supervision of physician or other certified health care professional Model: Z-800WF
When utlizing the patient query feature on the Zyno Medical Z-800WF pumps with software version 5.2.05, the alarm volume may inadvertently revert from a higher volume setting to a low one, causing potential delays in therapy and in extreme cases, associated risk of organ failure or death