Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2025-10-01FDA-DeviceBiomet, Inc.Class IIJuggerknot Mini Soft Anchors, Model/Catalog Number: 912080
The outer carton box label is for the JuggerKnot 1.0mm Mini products. However, the products within the carton are JuggerKnot 1.4mm Short devices.
- 2025-10-01FDA-DeviceBoston Scientific CorporationClass IIExtractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 (9-12mm Above);
The product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and vice versa.
- 2025-10-01FDA-DeviceORTOMA ABClass IIOrtoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201 Model Number: 2001 UDI-DI code: 07350137521019 Inserter Adapter OTD Pinnacle Straight REF 30-202 Model Number: 1001 UDI-DI code: 07350137520593 Reamer Adapter OTD REF 30-157 Model Number: 1003 UDI-DI code: 07350137520081 Inserter Adapter OTD REF 30-158 Model Number: 1004 UDI-DI code: 07350137520098 The Inserter Adapters are intended to be attached to a Cup Inserter Tool. The Cup Inserter Tool (impactor) is
Due to a weld breakage between the pin and inserter holder of the Inserter Adapter.
- 2025-10-01FDA-DeviceBiofire DefenseClass IIFilmArray NGDS Warrior Panel NGDS-ASY-0007 is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system.
FilmArray NGDS Warrior Panel (NGDS-ASY-0007) has an increased risk of internal control failures when testing positive blood culture, which may lead to delayed results and additional diagnostic workup.
- 2025-10-01FDA-DeviceCooperVision, Inc.Class IIVoyant 1-Day Premium Toric, (stenfilcon A) contact lens. Model Number: Voyant 1-Day Premium Toric. 54% water content, Toric, cast moulded Silicone Hydrogel, In-monomer tinted, contact lens, for Daily Disposable Wear. Blister packaged.
One lot manufactured with an invalid sterilization cycle.
- 2025-10-01FDA-DeviceCooperVision, Inc.Class IISofmed Breathables, (stenfilcon A) contact lens. Model Number: Softmed Breathables. 54% water content, Toric, cast moulded Silicone Hydrogel, In-monomer tinted, contact lens, for Daily Disposable Wear. Blister packaged.
One lot manufactured with an invalid sterilization cycle.
- 2025-10-01FDA-DeviceCooperVision, Inc.Class IIMyDay Toric, stenfilcon A) Contact Lens. Model/Catalog Number: MyDay Toric. 54% water content, Toric, cast moulded Silicone Hydrogel, In-monomer tinted, contact lens, for Daily Disposable Wear. Blister packaged.
One lot manufactured with an invalid sterilization cycle.
- 2025-10-01FDA-DeviceETHICON, LLCClass IISTRATAFIX Spiral PDS Plus Bidirectional Knotless Tissue Control Device, SFX SPI PDS+ BI VIO 14X14IN(36X36CM) USP1(M4) D/A MO-4. Product Code: SXPP2B400-11.
Potential for barb non-engagement.
- 2025-10-01FDA-DevicePhilips North America LlcClass IIPhilips CT systems labeled as the following: 1. CT 6000, Model Number: 728307.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- 2025-10-01FDA-DevicePhilips North America LlcClass IIPhilips CT systems labeled as the following: 1. Ingenuity CT Family Brazil, Model Number: 728325.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- 2025-10-01FDA-DevicePhilips North America LlcClass IIPhilips CT systems labeled as the following: 1. Ingenuity Core128 / Elite China, Model Number: 728324.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- 2025-10-01FDA-DevicePhilips North America LlcClass IIPhilips CT systems labeled as the following: 1. IQon Spectral CT, Model Number: 728332.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- 2025-10-01FDA-DevicePhilips North America LlcClass IIPhilips CT systems labeled as the following: 1. Ingenuity CT 728326, Model Number: 728321. 2. Ingenuity CT 728326, Model Number: 728323. 3. Ingenuity CT, Model Number: 728326.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- 2025-10-01FDA-DevicePhilips North America LlcClass IIPhilips CT systems labeled as the following: 1. Brilliance iCT SP, Model Number: 728311.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- 2025-10-01FDA-DevicePhilips North America LlcClass IIPhilips CT systems labeled as the following: 1. Brilliance iCT, Model Number: 728306.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- 2025-10-01FDA-DevicePhilips North America LlcClass IIPhilips CT systems labeled as the following: 1. Big Bore RT, Model Number: 728242. 2. Brilliance CT Big Bore Oncology, Model Number: 728243. 3. Brilliance CT Big Bore, Model Number: 728244.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- 2025-10-01FDA-DevicePhilips North America LlcClass IIPhilips CT systems labeled as the following: 1. Brilliance 16 Power, Model Number: 728240. 2. Brilliance CT 16 Slice, Model Number: 728246. 3. Brilliance CT 6 Slice, Model Number: 728256.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- 2025-10-01FDA-DevicePhilips North America LlcClass IIPhilips CT systems labeled as the following: 1. Brilliance CT 40 Channel, Model Number: 728235. 2. Brilliance CT 64 Channel, Model Number: 728231. 3. Ingenuity Flex, Model Number: 728317
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- 2025-10-01FDA-DeviceBaxter Healthcare CorporationClass IIBaster Continu-Flo Solution Set, Non-Vented, 2 Clearlink Luer Activated Valves, Backcheck Valve. 10 drops/mL. 112" (2.8 m), Product Code REF 2C8519; intravascular administration set
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for one lot of Continu-Flo solution sets listed below due to customer reports of tubing separation.
- 2025-10-01FDA-DeviceAbiomed, Inc.Class IAutomated Impella Controller (AIC) with the below brands and product codes. Used with heart assist pump. 1. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 2. Impella Controller, Packaged, US, Loaner; Product Code: 0042-0000-US-L. 3. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 4. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 5. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 6. Impella Controller, UK, Loaner; Product Code: 0
Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.
- 2025-10-01FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IMedline medical convenience kits labeled as: 1. NEWBORN RESUSCITATION KIT, Kit Number ACC010381C; 2. DRAWER 5 - POSITION 3, Kit Number ACC010532; 3. DRAWER #4C ADULT, Kit Number ACC010671; 4. DRAWER #2, Kit Number ACC010728.
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles due to the failure of the light to illuminate as intended.
- 2025-10-01FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IMedline medical convenience kits labeled as: 1. ADULT INTUBATION TRAY, Kit Number ACC010369B; 2. ADULT ICU INTUBATION TRAY, Kit Number ACC010379A; 3. INTUBATION TRAY, Kit Number ACC010392; 4. INTUBATION KIT, Kit Number ACC010467A; 5. ADULT AIRWAY TRAY, Kit Number ACC010469; 6. TRAY 2 INTUBATION DRAWER, Kit Number ACC010480; 7. TRAY 2 INTUBATION DRAWER, Kit Number ACC010480A; 8. ADULT INTUBATION, Kit Number ACC010540; 9. DIFFICULT AIRWAY TRAY #1, Kit Number ACC010593A; 10. DI
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles due to the failure of the light to illuminate as intended.
- 2025-09-24FDA-DeviceC.R. Bard IncClass IISureStep Foley Tray System BARD¿ Lubri-Sil Foley Catheter Tray, REF: A942216
Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection control, when it's silicone hydrogel coated without antimicrobial coating, which may cause infection and/or local and/or systemic allergic reaction
- 2025-09-24FDA-DeviceC.R. Bard IncClass IISureStep Foley Tray System Lubri-Sil I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303414A
Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection control, when it's silicone hydrogel coated without antimicrobial coating, which may cause infection and/or local and/or systemic allergic reaction
- 2025-09-24FDA-DeviceC.R. Bard IncClass IISureStep Foley Tray System Bardex I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303316A
Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection control, when it's silicone hydrogel coated without antimicrobial coating, which may cause infection and/or local and/or systemic allergic reaction
- 2025-09-24FDA-DeviceMaquet Cardiovascular, LLCClass IIHeartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
- 2025-09-24FDA-DeviceMaquet Cardiovascular, LLCClass IIHeartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
- 2025-09-24FDA-DeviceMaquet Cardiovascular, LLCClass IIHeartstring III Proximal Seal System. Intravascular anastomosis occluder.
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
- 2025-09-24FDA-DeviceAbbott Point Of Care Inc.Class IIi-STAT EG6+ cartridge. List Number: 03P77-25.
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.
- 2025-09-24FDA-DeviceAbbott Point Of Care Inc.Class IIi-STAT EG7+ cartridge. List Number: 03P76-25.
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.