- 2025-10-08FDA-DeviceNortheast Scientific Inc.Class IINES Reprocessed 1.4mm RX Turbo Elite Laser Atherectomy Catheter. Model Number: R-414-159. The NES Reprocessed Turbo-Elite Laser Atherectomy Catheter is indicated for use in the treatment, including atherectomy, of infrainguinal stenoses and occlusions. This model is not cleared for marketing in the US.
Potential for breaches in the sterile barrier packaging, compromising sterility assurance.
- 2025-10-08FDA-DeviceNortheast Scientific Inc.Class IINES Reprocessed 0.9mm RX Turbo Elite Laser Atherectomy Catheter. Model/Catalog Number: R-410-154. The NES Reprocessed Turbo-Elite Laser Atherectomy Catheter is indicated for use in the treatment, including atherectomy, of infrainguinal stenoses and occlusions. This model is not cleared for marketing in the US.
Potential for breaches in the sterile barrier packaging, compromising sterility assurance.
- 2025-10-08FDA-DeviceNortheast Scientific Inc.Class IINES Reprocessed 2.0mm OTW Turbo Elite Laser Atherectomy Catheter. Model Number: R-420-006. The NES Reprocessed Turbo-Elite Laser Atherectomy Catheter is indicated for use in the treatment, including atherectomy, of infrainguinal stenoses and occlusions. This model is not cleared for marketing in the US.
Potential for breaches in the sterile barrier packaging, compromising sterility assurance.
- 2025-10-08FDA-DeviceNortheast Scientific Inc.Class IINES Reprocessed 1.7mm OTW Turbo Elite Laser Atherectomy Catheter. Model Number: R-417-152. The NES Reprocessed Turbo-Elite Laser Atherectomy Catheter is indicated for use in the treatment, including atherectomy, of infrainguinal stenoses and occlusions. This model is not cleared for marketing in the US.
Potential for breaches in the sterile barrier packaging, compromising sterility assurance.
- 2025-10-08FDA-DeviceNortheast Scientific Inc.Class IINES Reprocessed 1.4mm OTW Turbo Elite Laser Atherectomy Catheter. Model Number: R-414-151. The NES Reprocessed Turbo-Elite Laser Atherectomy Catheter is indicated for use in the treatment, including atherectomy, of infrainguinal stenoses and occlusions. This model is not cleared for marketing in the US.
Potential for breaches in the sterile barrier packaging, compromising sterility assurance.
- 2025-10-08FDA-DeviceNortheast Scientific Inc.Class IINES Reprocessed 0.9mm OTW Turbo Elite Laser Atherectomy Catheter. Model Number: R-410-152. The NES Reprocessed Turbo-Elite Laser Atherectomy Catheter is indicated for use in the treatment, including atherectomy, of infrainguinal stenoses and occlusions.
Potential for breaches in the sterile barrier packaging, compromising sterility assurance.
- 2025-10-08FDA-DeviceS.I.N. Implant System LtdaClass IIIMPLANT EPIKUT PLUS MT16 3,5X10,0MM, REF: ILM 3510N; IMPLANT EPIKUT PLUS MT16 3.8X11.5MM, REF: ILM 3511N; IMPLANT EPIKUT PLUS MT16 3,5X13,0MM, REF: ILM 3513N; IMPLANT EPIKUT PLUS MT16 3,5X15,0MM, REF: ILM 3515N; IMPLANT EPIKUT PLUS MT16 3,8X10,0MM, REF: ILM 3810N; IMPLANT CM16 3,8 X 11MM, REF: ILM 3811N; IMPLANT EPIKUT PLUS MT16 3,8X15,0MM, REF: ILM 3815N; IMPLANT EPIKUT PLUS MT16 4,0X10,0MM, REF: ILM 4010N; IMPLANT EPIKUT PLUS MT16 4,0X11,5MM, REF: ILM 4011N; IMPLANT EPIKUT PLUS
Dental implants were imported under temperature and storage conditions that do not comply with the instructions for use and product labeling so there may be packaging damage, which could compromise product quality and performance.
- 2025-10-08FDA-DeviceGetinge Disinfection AbClass IIGetinge 88-Series Washer- Disinfector. Model Number: 88-5.
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IICardio Vascular-Allura Centron; Model Numbers: 722400; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper CV20; Model Numbers: 722031; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20/20 OR Table; Model Numbers: 722039; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20/20; Model Numbers: 722038; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20/15 OR Table; Model Numbers: 722059; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20/15; Model Numbers: 722058; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20/10; Model Numbers: 722029; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20 OR Table; Model Numbers: (1) 722015, (2) 722023, (3) 722035; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20 Biplane OR Table; Model Numbers: (1) 722020, (2) 722025; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20 Biplane; Model Numbers: (1) 722008, (2) 722013; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20; Model Numbers: (1) 722006, (2) 722012, (3) 722028; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD10 OR Table; Model Numbers: 722033; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD10/10; Model Numbers: (1) 722005, (2) 722011, (3) 722027; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IICheck your Philips Allura Xper FD10 system
Certain Philips Allura Xper FD10 X-ray systems may have a BIOS battery that depletes faster than expected, preventing the system from starting up. The system will not display any warning before the battery runs out of power.
- 2025-10-08FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIContact Philips about Allura Xper FD10C system failure
The BIOS battery in some Philips Allura Xper FD10C X-ray systems may run out of power faster than expected. When this happens, the system will not start up, and there are no warning messages to alert users before the battery fails.
- 2025-10-01FDA-DeviceMaquet Cardiovascular, LLCClass IIVasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.
Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
- 2025-10-01FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura CV20; Product Code: 722031; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- 2025-10-01FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20/15 OR Table; Product Code: 722059; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- 2025-10-01FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20/15; Product Code: 722058; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- 2025-10-01FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20/10; Product Code: 722029; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- 2025-10-01FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20/20 OR Table; Product Code: 722039; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- 2025-10-01FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD20/20; Product Code: 722038; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.