Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2025-05-07FDA-DeviceSiemens Healthcare Diagnostics IncClass IIepoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736382; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.
The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
- 2025-05-07FDA-DeviceSiemens Healthcare Diagnostics IncClass IIepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.
The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
- 2025-05-07FDA-DeviceLeica Biosystems Melbourne Pty LtdClass IIHistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4.0. The HistoCore PELORIS 3 Rapid Tissue Processor is a dual retort rapid tissue processor used to prepare tissue samples.
There is a leakage issue associated with the tubing in the manifold of the instrument.
- 2025-05-07FDA-DeviceBelmont Instrument LLCClass II3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume Reservoirs (P/N 903-00018) and 4 Heat Exchanger/Patient Line set (P/N 403-00382). The two components are packaged separately in individual pouches and placed in the same case for shipment. The female quick connector that may be cracked is located on the Heat Exchanger/Patient Line set.
Potential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during priming of the Rapid Infuser.
- 2025-05-07FDA-DeviceBelmont Instrument LLCClass IIStraight Inflow/Bifurcated Outflow Line Kit. Model Number: 902-00038. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient Line Kit (P/N 902-00038) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and Straight Inflow/Bifurcated Outflow Patient Line (P/N 902-00040P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three component
Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.
- 2025-05-07FDA-DeviceBelmont Instrument LLCClass IIStraight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient Line Kit (P/N 902-00037) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and Straight Inflow/Outflow Patient Line (P/N 902-00039P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three components are connecte
Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.
- 2025-05-07FDA-DeviceBelmont Instrument LLCClass IIHyperthermia Pump Procedure Kit. Model Number: 902-00045. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Hyperthermia Procedure kit (P/N 902-00045) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and HP Procedure Pack (P/N 902-00048P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three components are connected prior to use. The female quick connector
Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.
- 2025-05-07FDA-DeviceSiemens Healthcare Diagnostics IncClass IIStratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I in heparinized plasma.
Siemens Healthcare Diagnostics Inc., is recalling their Stratus CS Acute Care cTNI TestPak, an IVD test for cardiac troponin I in heparinized plasma by removal. The reason for the recall is an increased occurrence of random non-repeatable false positive cardiac Troponin l(cTnI) results at any point during the TestPaks shelf life when using the Stratus CS cTnI Acute Care Testpak.
- 2025-05-07FDA-DeviceBeckman Coulter Inc.Class IIDxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065 Software versions: V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3 The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
The reason for the recall is incorrect sample barcode identification and processing errors that result in the reporting of erroneous and delayed results.
- 2025-05-07FDA-DeviceTrimble AbClass IIThe Trimble RTS873 robotic total station is a solution designed for construction professionals seeking precision, efficiency, and reliability in their layout tasks. The integrated Class 2 green aiming laser system enhances visibility and accuracy, even on uneven surfaces. It is ideal for general and specialty contractors, the RTS873 streamlines layout processes, reduces rework, and enhances overall productivity.
An issue has been identified in the production of the RTS873 robotic total stations that may result in the green aiming laser output power being above its laser safety Class 2 classification.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare CARESTATION 750c A2, Model/REF Number 1012-9755-002
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare CARESTATION 750c A1, Model/REF Number 1012-9755-000
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare CARESTATION 750 A2, Model/REF Number 1012-9750-002
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare CARESTATION 750 A1, Model/REF Number 1012-9750-000
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare Carestation 650c A1, Model/REF Number 1012-9655-200
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare CARESTATION 650C A2, Model/REF Number 1012-9655-002
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare Carestation 650 A1, Model/REF Number 1012-9650-200
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare Carestation 650 SE, Model/REF Number 1012-9650-012
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare CARESTATION 650 A2, Model/REF Number 1012-9650-002
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare CARESTATION 650 A1, Model/REF Number 1012-9650-000
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare Carestation 620 A1, Model/REF Number 1012-9620-200
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare Carestation 620 SE, Model/REF Number 1012-9620-012
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare CARESTATION 620 A2, Model/REF Number 1012-9620-002
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceGE Medical Systems China Co., Ltd.Class IGE Healthcare CARESTATION 620 A1, Model/REF Number 1012-9620-000
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-04-30FDA-DeviceStraumann USA LLCClass IIWB XL Anatomic Healing Abutment XC. Endosseous Dental Implant Abutments.
Incorrect blister labelling.
- 2025-04-30FDA-DeviceMicro-X Ltd.Class IIBrand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number:MXU-RV35 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorporates a self-contained X-ray generator, imaging display and software for acquiring medical diagnostic images outside a standard X-ray room. It is designed specifically for Digital Radiography (DR) with vendor-supported detectors. The Rover is designed to perform radiographic X-ray examinations for disease/injury de
Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
- 2025-04-30FDA-DeviceMicro-X Ltd.Class IIBrand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71, MXU-RV35 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorporates a self-contained X-ray generator, imaging display and software for acquiring medical diagnostic images outside a standard X-ray room. It is designed specifically for Digital Radiography (DR) with vendor-supported detectors. The Rover is designed to perform radiographic X-ray examinations for diseas
Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
- 2025-04-30FDA-DeviceRemote Diagnostic Technologies Ltd.Class IITempus Pro Patient Monitor, REFs: 00-1004-R (Tempus Pro non-printer, IBP, without headset dock), 00-1007-R and 00-1007 (Tempus Pro non-printer, IBP, with headset dock), 00-1024-R (Tempus Pro printer), 00-1026-R (Tempus Pro printer, IBP), 00-1007 [(Tempus Pro, BT Dock (IB Only)], 00-1004 00-1004 [Tempus Pro (IB Only)]
software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.
- 2025-04-30FDA-DeviceAbbott Molecular, Inc.Class IIThe Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.