FDA-Device2026-10-14Class IIPROCESSING DEFECT

Steris VERIFY Biological Integrator Challenge Pack recalled for inadequate spore population

OTHERNationwide distributionMedical devices

What this means

Some Steris VERIFY Biological/Integrator Challenge Packs (Model LCB006) may contain fewer sterilization-testing spores than specified. This could cause the test to show a successful steam sterilization cycle when the cycle did not actually kill enough microorganisms to be safe.

  • Check if you have lots 260810, 260810A, 261106, or 270116 with expiration dates between August 2026 and January 2027.
  • Contact Steris Corporation or your supplier for instructions on affected units and replacements.
  • Do not use affected packs to validate steam sterilization cycles until you hear from Steris.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Intended for
All ages
Illnesses reported
None reported
Sold states
Worldwide - US Nationwide distribution.
Affected count
266
Manufactured in
5960 Heisley Rd, N/A, Mentor, OH, United States
Products
VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0070-2027