FDA medical device recalls
Formal FDA enforcement reports for medical devices — implants, monitors, diagnostic tests, and hospital and home-use equipment. 961 recalls in the last 12 months, in plain English.
Recent recalls
Page 1 of 33- Intelerad Ambra PACS recalled for incomplete study risk
Certain versions of Ambra PACS (InteleShare) imaging software have two issues: studies may be duplicated incompletely, and images may be archived incompletely. A reader could act on an incomplete examination, which could lead to misdiagnosis, delayed diagnosis or treatment, or the need for additional imaging.
ModerateFDANov 20, 2026Intelerad
- Intelerad IntelePACS software recalled for risk of deleting studies
Intelerad Medical Systems is recalling IntelePACS radiological image processing software versions PACS-5-2-1 through PACS-5-9-1 because a software issue may delete studies outside of the defined deletion criteria. The recall covers 39 units distributed across 14 states. The notice reports no injuries.
ModerateFDANov 20, 2026Intelerad
- 3-D Matrix PuraGel recalled for incomplete instructions for use
3-D Matrix is recalling PuraGel because its instructions for use are missing additions that FDA requested during the final review of the premarket submission. No injuries have been reported. The notice covers two sizes, 1 mL (reference 651-004) and 3 mL (reference 651-005).
ModerateFDANov 20, 20263-D Matrix
- Paragon 28 Grappler suture anchors recalled for possible tension loss
Paragon 28, Inc. is recalling R3FLEX Stabilization System Grappler Suture Anchor Systems because they lack a pre-tied knot. This could mean the anchor assembly does not properly hold tension. No injuries have been reported.
ModerateFDANov 20, 2026Paragon 28
- Medline Bubble CPAP kit recalled for possible heat damage risk
Medline Industries is recalling Medline Bubble CPAP kits because neonatal and infant heated wire breathing circuits may fail sporadically. The heated wire connector at the humidifier may show discoloration, melting, smoke, burning odor, or other signs of localized heat damage. The notice reports no injuries, and 24 units were distributed in Florida and Texas.
ModerateFDANov 20, 2026Medline
- DePuy Attune femoral knee implant recalled for mislabeled size
DePuy Orthopaedics is recalling Attune CR cemented femoral knee implants, size 5N right, product number 150400225, because size 4 implants may be labeled as size 5N on the outer boxes. If the mislabeling is found before implantation, it may cause a minor surgical delay. Implanting an oversized device may lead to bone damage, loosening, or poor joint mechanics.
ModerateFDANov 20, 2026DePuy Orthopaedics
- Hologic Selenia and 3Dimensions mammography systems recalled for freefall risk
Hologic is recalling two mammography system lines, Selenia Dimensions and 3Dimensions System 2D International, for a potential risk of sudden, unexpected freefall of the C-Arm. The risk comes from significant wear on the vertical lift nut of the C-Arm assembly. The notice lists 16,089 affected Selenia Dimensions units and 7,115 affected 3Dimensions units, and reports no injuries.
ModerateFDANov 20, 2026Hologic2 products
- Medline surgical procedure kits recalled for reprocessed components
Medline is recalling 163 medical procedure kits (48 and 115 units in two product groups) that contain reprocessed components recalled after the kits were distributed. The kits were sold as Medline Kit SKUs for laparoscopic colorectal, lap chole, GYN/LAVH and robotic general procedures, and were distributed nationwide, with units in MD, NY and OK. The FDA classified this as a Class II recall, and no injuries have been reported.
ModerateFDANov 20, 2026Medline2 products
- Coloplast Aris Transobturator Kit recalled for mislabeled expiration
Coloplast's Aris Transobturator Kit, catalog number 93-4400, has an outer box expiration date that is later than the expiration date on the individually packaged mesh inside. The notice covers lots 222-08273991 and 222-08556313, with 167 units affected and US nationwide distribution. No injuries have been reported.
ModerateFDANov 20, 2026Coloplast
- Diversatek SafeGuide guidewires recalled for unexpected flexibility
Diversatek Healthcare is recalling SafeGuide Marked Spring Tip Guidewires from affected lots because they are more flexible than expected. The recall covers 228 units of model 1214-02 (catalog PSN421V), distributed nationwide in the US. No injuries have been reported.
ModerateFDANov 20, 2026Diversatek Healthcare2 products
- Medtronic Abre venous stent system recalled for partial deployment risk
Medtronic is recalling 62 catalog numbers of the Abre Venous Self-Expanding Stent System after an observed increase in partial stent deployments. The notice says partial deployment is associated with potential harms, including longer procedures, a secondary procedure or surgery, embolism, blocked blood flow, vessel perforation, rupture, and blood loss. No injuries have been reported.
ModerateFDANov 20, 2026Medtronic
- Boston Scientific OptiCross catheters need new instructions for air embolism
Boston Scientific is updating the instructions for use for six OptiCross coronary and peripheral imaging catheter lines, covering about 1.29 million units. The update follows reports of air embolism events and of repeated flushing to restore image quality while the catheter was inside the patient. The agency says intravascular air may cause air embolism, with effects of varying severity such as heart rhythm and blood flow problems, heart attack, cardiac arrest, and other serious injury. No injuries are reported in the notice.
SeriousFDANov 20, 2026Boston Scientific6 products
- Saluda Evoke stimulator recalled for short circuit risk
Saluda Medical is recalling Evoke spinal cord stimulators, models 3042 and 1042, because they may have short circuits. The short circuits could prevent the device from charging or cause unintended stimulation during charging. No injuries have been reported.
ModerateFDANov 20, 2026Saluda Medical
- SKYTRON INTEGRITY 215 SG sterilizer recalled for overheating risk
Sakura Seiki Co., Ltd. is recalling 81 SKYTRON INTEGRITY Model 215 SG steam sterilizers because they may overheat and shut down. The FDA classifies this as a Class II recall, and no injuries have been reported. The units were distributed in 20 states, including CA, FL, IL, OH, and TX.
ModerateFDANov 20, 2026Sakura Seiki
- Maine Molecular FilmArray GI Control Panel recalled for mislabeling
Some FilmArray GI Control Panel M238 kits, quality control products used with BIOFIRE GI Panel assays, are labeled "Research Use Only." The affected kits were distributed in the US and Germany. No injuries have been reported.
ModerateFDANov 20, 2026Maine Molecular Quality Controls
- Abbott i-STAT hs-TnI cartridges recalled for possible false troponin results
Abbott Point of Care is recalling i-STAT hs-TnI cartridges because a component used in reagent manufacturing shows variability. This may cause falsely elevated cardiac troponin I results, which may lead to unnecessary clinical intervention, such as cardiac catheterization without true evidence of myocardial injury. No injuries have been reported.
ModerateFDANov 20, 2026Abbott
- RayStation software recalled for potential dose deformation
RaySearch Laboratories is recalling 9 product entries of its RayStation radiation therapy treatment planning software, from RayStation 6.1.1 through RayStation 2024A SP3, distributed nationwide in the US. The agency says a deformable registration that expands outward near the boundary of the original dose grid may deform the dose, possibly causing the clinically relevant dose to be lost. The notice reports no injuries.
ModerateFDANov 18, 2026Raysearch Laboratories9 products
- Philips Azurion systems recalled for restart power supply failure
When a power supply failure occurs during a start or restart, the Power Distribution Unit does not provide the required direct current power to essential system components. The system does not complete the restart, imaging is unavailable, and the system remains unavailable for clinical use. The recall covers 2,632 units across eight Azurion 3, 5 and 7 model numbers.
ModerateFDANov 18, 2026Philips
- Roche cobas c 303 analyzer recalled for possible erroneous results
Roche Diagnostics is recalling cobas c 303 analytical units because vacuum tubing in the cell rinse unit may deteriorate early from excess bending stress. The tubing may develop pinholes, cracks or ruptures, and fluid may drop into reaction cells without any system alarm, which could lead to erroneous patient and quality control results.
ModerateFDANov 18, 2026Roche Diagnostics
- B Braun Streamline bloodline sets recalled for possible leaks
B Braun is recalling Streamline Bloodline Sets for Dialog+ (material number SL-2010M2096) because the arterial tubing connection to the dialyzer connector may not be bonded properly. This could cause the tubing to detach, separate, or leak during priming or dialysis treatment. The notice reports no injuries.
SeriousFDANov 18, 2026B. Braun
- Philips EPIQ ultrasound systems recalled for possible control failure
Philips Ultrasound, LLC is recalling EPIQ ultrasound systems because their TGC slide controls, which adjust gain compensation over a 2D image region, may become unresponsive after multiple start-ups or reboots. Rebooting may not restore the controls. The notice reports no injuries.
ModerateFDANov 17, 2026Philips
- Beckman Coulter LabPro V5.0 update recalled for software defects
Beckman Coulter is recalling MicroScan LabPro V5.0 Maintenance Update 1 with Panel Update 09 kits because of software defects that may affect antimicrobial susceptibility testing. The notice reports no injuries. The product was distributed internationally.
ModerateFDANov 17, 2026Beckman Coulter
- Summit Medical bone cement mixing bowl recalled for possible endotoxin
Summit Medical Ltd. is recalling vacuum mixing bowls used with PMMA bone cement. Scraper components in the bowls may have endotoxin levels outside specification because of a processing error. No injuries have been reported.
ModerateFDANov 17, 2026Summit Medical
- Philips EPIQ Elite ultrasound recalled for possible misdiagnosis risk
A software issue in the Philips EPIQ Elite Diagnostic Ultrasound System, Model 795125, can cause the Doppler sample to display shallower than its actual measurement location. This could lead to inaccurate blood flow velocity measurements and possible misdiagnosis. The notice reports no injuries.
ModerateFDANov 17, 2026Philips
- Helena V8 SPE kit recalled for possible albumin error
The V8 SPE Diluent in this serum protein kit may lower the albumin result and raise the beta and gamma globulin results. Quality control may fail, so no results may be reported and a repeat sample may be needed. An incorrect result could lead to unnecessary clinical workup or hide a mild true elevation in albumin.
ModerateFDANov 17, 2026Helena Laboratories
- Physio-Control LIFEPAK 35 monitors recalled for warning display risk
Physio-Control, Inc. is recalling certain LIFEPAK 35 monitor/defibrillators. The notice says the devices may not display warning messages as intended, so users may not receive expected visual alerts, which could delay treatment. No injuries have been reported.
ModerateFDANov 17, 2026Physio-Control
- O&M Halyard laparoscopy pack recalled for inconsistent latex-free labeling
O&M Halyard Inc. is recalling Laparoscopy Comprehensive Pack II, model 49332, for inconsistent latex-free labeling. The pack contains a component made with natural rubber latex, but its labeling says it is not made with natural rubber latex. The notice reports 396 units distributed in California and Michigan and no injuries.
ModerateFDANov 17, 2026Halyard
- Cardinal Health Presource kits recalled for possible tubing leaks
Cardinal Health is recalling 18 Presource Convenience procedure kits that contain WelMed Extension Tubing Sets, which WelMed recalled from its supplier chain. The affected tubing sets may leak. No injuries have been reported, and the kits were distributed in the US and Canada.
ModerateFDANov 13, 2026Cardinal Health18 products
- GE Healthcare SIGNA Artist MRI recalled for potential arc flash
GE Healthcare is recalling 1.5T SIGNA Artist Evo UPGRADE and SIGNA Artist FLO IPM MRI systems. An exposed sharp edge inside an electronic cabinet could damage power cables, which could potentially cause an electrical fault such as an arc flash. No injuries have been reported.
ModerateFDANov 13, 2026GE Healthcare
- Wright Medical EVOLVE TRIAD drill guide recalled for fracture risk
Wright Medical Technology is recalling EVOLVE TRIAD drill guides, reference 49510109, that may not have been manufactured to specification. The drill guide may fracture under less force than intended during use. The notice reports no injuries, and 62 units are affected.
ModerateFDANov 13, 2026Wright Medical