FDA-Device2026-10-14Class IIPROCESSING DEFECT

Steris VERIFY Biological Steam Test Pack recalled for low spore population

OTHERNationwide distributionMedical devices

What this means

Steris Corporation's VERIFY Biological Steam Test Pack (Model S3065) may contain fewer bacterial spores than required. This could cause the test to incorrectly show that steam sterilization worked properly, even when it did not adequately kill harmful microorganisms on medical equipment.

  • Check if you have lot numbers 260810, 260810A, 261106, or 270116 with the expiration dates shown
  • Contact Steris Corporation for guidance on affected test packs
  • Do not rely on test results from affected lots to confirm sterilization is complete
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Intended for
All ages
Illnesses reported
None reported
Sold states
Worldwide - US Nationwide distribution.
Affected count
217
Manufactured in
5960 Heisley Rd, N/A, Mentor, OH, United States
Products
VERIFY Biological Steam Test Pack Model/Catalog Number: S3065

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0074-2027