FDA-Device2026-10-20Class IPROCESSING DEFECT

Ventec Life Systems VOCSN ventilator recalled for potential ventilation interruption

OTHERNationwide distributionMedical devices

What this means

The VOCSN ventilator's internal flow sensor cable connector may disconnect from the device's internal board. If this happens, the ventilator may restart unexpectedly and temporarily stop delivering breathing support.

  • Contact Ventec Life Systems immediately if you own or operate a VOCSN ventilator (Model PRT-01100-000)
  • Do not rely solely on this device until you receive further guidance from the manufacturer
  • Have a backup ventilation plan in place and consult your clinician right away
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

Intended for
Adults
Illnesses reported
None reported
Sold states
Distribution US nationwide including Puerto Rico.
Affected count
1,532 units
Manufactured in
22002 26th Ave Se, Bothell, Washington, United States
Products
VOCSN Model Number: PRT-01100-000, Package PN:PRT-01126-000. used for ventilatory support.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0126-2027