FDA-Device2026-10-21Class IIPROCESSING DEFECT

JenaValve Trilogy Delivery Catheter (25MM) recalled for sleeve liner separation risk

OTHERNationwide distributionMedical devices

What this means

JenaValve Trilogy Delivery Catheter (Size 25MM, REF: TRILOGY-DC-M) may experience separation of the sleeve liner, which can prevent or delay deployment of the heart valve device during surgery. This affects 108 units distributed across the US and several European countries.

  • If you have one of the affected lot numbers (DS-040826-005, DS-040926-002, DS-042326-001, DS-042826-007, DS-042926-008, DS-050826-005, DS-051526-001), do not use it.
  • Contact JenaValve Technology Inc. or your medical device supplier immediately for a replacement or corrective action.
  • Healthcare providers should verify lot numbers before any scheduled procedures.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.

Intended for
Adults
Illnesses reported
None reported
Sold states
US Nationwide distribution in the states of AZ, CA, FL, GA, IL, MA, MD, MI, MN, MO, NY, OH, PA, SC, TX, UT, VA, WA and the countries of Germany, Denmark, Great Britain, Italy.
Affected count
108 units
Manufactured in
4 Cromwell, Irvine, CA, United States
Products
JenaValve Trilogy Delivery Catheter, Size 25MM REF: TRILOGY-DC-M

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0135-2027

More Jenavalve recalls

Want us to check the next one for you?

We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Free · One email a week · Unsubscribe anytime
By signing up you agree to our Terms and Privacy Policy.