FDA-Device2026-10-22Class IIPROCESSING DEFECT
Boston Scientific Direxion Torqueable Microcatheter recalled for coating and manufacturing defects
What this means
Boston Scientific identified defects in certain Direxion Torqueable Microcatheters due to incomplete UV curing of the hydrophilic coating on the device tip and misaligned laser cuts that created raised edges on internal components. These defects could affect device performance during medical procedures.
- If you have one of the affected batch numbers, contact your hospital or surgical center immediately
- Do not use the device if you have confirmed it matches a recalled batch number
- Check the material number (M001195240) and batch number on your device
The agency's original wording
Hazard
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Intended for
Adults
Illnesses reported
None reported
Sold states
Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T¿rkiye, UAE, United Kingdom.
Affected count
8542
Manufactured in
Boston Scientific Limited, Business And Technology Park, Model Farm Road, Cork, Ireland
Products
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/1RO/130;
Material Number: M001195240
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0227-2027More Boston Scientific recalls
- FDA-Device2026-10-22Boston Scientific Direxion Torqueable Microcatheter recalled for coating and manufacturing defects
- FDA-Device2026-10-22Boston Scientific Direxion Torqueable Microcatheter recalled for coating and manufacturing defects
- FDA-Device2026-10-22Boston Scientific Direxion Torqueable Microcatheter recalled for coating and manufacturing defects
- FDA-Device2026-10-22Boston Scientific Direxion Torqueable Microcatheter recalled for coating and manufacturing defects
- FDA-Device2026-10-22Boston Scientific Direxion Torqueable Microcatheter recalled for coating and manufacturing defects
Want us to check the next one for you?
We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.