FDA-Device2026-10-22Class IIPROCESSING DEFECT

Boston Scientific Direxion Torqueable Microcatheter recalled for coating and manufacturing defects

OTHERNationwide distributionMedical devices

What this means

Boston Scientific is recalling certain Direxion Torqueable Microcatheters due to manufacturing defects: the hydrophilic coating on the catheter tip may not have been properly cured, and nitinol sleeves may have raised ridges from laser cutting misalignment. These defects could affect the catheter's performance during use.

  • Check if you have a device with Material Number M001195290 and one of these batch numbers: 38362172, 38362173, 38530450, or 38595659.
  • Contact Boston Scientific or your healthcare provider for instructions on device inspection or replacement.
  • Do not use the device if you suspect it has the defect.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

Intended for
Adults
Illnesses reported
None reported
Sold states
Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T¿rkiye, UAE, United Kingdom.
Affected count
42
Manufactured in
Boston Scientific Limited, Business And Technology Park, Model Farm Road, Cork, Ireland
Products
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/105; Material Number: M001195290

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0232-2027

More Boston Scientific recalls

Want us to check the next one for you?

We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Free · One email a week · Unsubscribe anytime
By signing up you agree to our Terms and Privacy Policy.