FDA-Device2026-10-22Class IIPROCESSING DEFECT

Boston Scientific Direxion Torqueable Microcatheter recalled for coating and manufacturing defects

OTHERNationwide distributionMedical devices

What this means

Boston Scientific's Direxion Torqueable Microcatheter has a manufacturing defect affecting the hydrophilic coating on the distal tip and nitinol sleeve, caused by insufficient UV curing and laser cutting misalignment. The defect may compromise the device's performance during medical procedures. This is a Class II recall affecting approximately 3,284 units distributed worldwide.

  • If you are a healthcare facility or provider, check your inventory against the affected batch numbers listed in the FDA notice.
  • Do not use affected units; contact Boston Scientific for instructions on device return or replacement.
  • Report any adverse events or complications to the FDA MedWatch program.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

Intended for
Adults
Illnesses reported
None reported
Sold states
Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T¿rkiye, UAE, United Kingdom.
Affected count
3284
Manufactured in
Boston Scientific Limited, Business And Technology Park, Model Farm Road, Cork, Ireland
Products
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/130; Material Number: M001195300

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0233-2027