FDA-Device2026-10-22Class IIPROCESSING DEFECT

Boston Scientific Direxion Torqueable Microcatheter recalled for coating and manufacturing defects

OTHERNationwide distributionMedical devices

What this means

Boston Scientific has recalled certain Direxion Torqueable Microcatheters due to manufacturing defects: the hydrophilic coating on the distal tip may not have been properly cured, and the nitinol sleeves may have raised ridges from laser cutting misalignment. These defects could affect device performance during use.

  • If you are a healthcare provider or facility, check your inventory against the affected batch numbers (37881100 through 40510769; see full list in the FDA notice).
  • Contact Boston Scientific if you have devices from affected batches; do not use them until advised.
  • Report any adverse events to the FDA MedWatch program or to Boston Scientific directly.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

Intended for
Adults
Illnesses reported
None reported
Sold states
Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T¿rkiye, UAE, United Kingdom.
Affected count
1442
Manufactured in
Boston Scientific Limited, Business And Technology Park, Model Farm Road, Cork, Ireland
Products
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/2RO/130; Material Number: M001195670

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0254-2027

More Boston Scientific recalls

Want us to check the next one for you?

We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Free · One email a week · Unsubscribe anytime
By signing up you agree to our Terms and Privacy Policy.