FDA-Device2026-10-27Class IIPROCESSING DEFECT

BioMerieux EMAG system recalled over software error delaying results

OTHERNationwide distributionMedical devices

What this means

BioMerieux is recalling a limited number of EMAG nucleic acid sample preparation instruments used in clinical labs. After updating to firmware SP4 Patch 2, a random software error (Error 10003, Heater timeout) can cause delayed or no results when analyzing clinical samples. No injuries have been reported.

  • Check whether your EMAG system's serial number is on the affected list in the recall notice.
  • Contact BioMerieux for guidance on affected instruments before relying on their results.
  • If you see Error 10003, expect possible delays and follow your lab's backup testing plan.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2

Intended for
All ages (under 100 years)
Illnesses reported
None reported
Sold states
Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany, Austria.
Affected count
33
Manufactured in
Chemin De L'Orme, Marcy L'Etoile, France
Products
EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0301-2027