Serious riskFDA DeviceClass I

B Braun Streamline bloodline sets recalled for possible leaks

B Braun is recalling Streamline Bloodline Sets for Dialog+ (material number SL-2010M2096) because the arterial tubing connection to the dialyzer connector may not be bonded properly. This could cause the tubing to detach, separate, or leak during priming or dialysis treatment. The notice reports no injuries.

From the FDA notice · Read the official notice ↗ · All B. Braun recalls

What to do

  1. Dialysis providers should stop using affected lots: A2600099, A2600128, and A2600136.
  2. Check lot numbers and expiration dates (07JAN2029 or 07FEB2029) against the product. UDI-DI primary 04046964367786 and unit 04046964367779 identify the product.
  3. The notice doesn't say how to get a replacement or refund; contact B Braun Medical Inc.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Hemodialysis Bloodlines STREAMLINE BLOODLINE SET FOR DIALOG+. Material Number: SL-2010M2096. STREAMLINE BLOODLINE SET FOR DIALOG
Sold in
Nationwide
Made for
Adults (0 months and older)

Details

Units affected
66,262 units
Illnesses reported
None reported
Manufactured in
824 12th Ave, Bethlehem, Pennsylvania, United States
Classification
Class I
Recall number
Z-0313-2027

In the FDA's words

Description
Hemodialysis Bloodlines STREAMLINE BLOODLINE SET FOR DIALOG+. Material Number: SL-2010M2096. STREAMLINE BLOODLINE SET FOR DIALOG Potential inadequate bonding of the arterial tubing connection to the dialyzer connector may result in detachment, separation, or leaking of the arterial tubing during priming or dialysis treatment. Material Number: SL-2010M2096; UDI-DI Primary: 04046964367786; UDI-DI Unit: 04046964367779; Lot Number (Expiration date): A2600099 (07JAN2029), A2600128 (Exp 07JAN2029), A2600136 (07FEB2029). Distribution: Nationwide distribution in the US. Affected quantity: 66,262 units
Hazard
Potential inadequate bonding of the arterial tubing connection to the dialyzer connector may result in detachment, separation, or leaking of the arterial tubing during priming or dialysis treatment.