Moderate riskFDA DeviceClass II

Accriva Hemochron Jr. APTT cuvettes recalled for inaccurate results

Accriva Diagnostics is recalling certain Hemochron Jr. APTT test cuvettes, used to check low-dose heparin levels in blood. The clot-activating reagent inside some cuvettes may be out of position or missing, which can cause error codes or falsely high results. Falsely high results could lead to a lower anticoagulant dose than needed, which could raise the risk of blood clots. No injuries have been reported.

From the FDA notice · Read the official notice ↗ · All Accriva Diagnostics recalls

What to do

  1. Check cuvette lot numbers against the recall list: B6JCA005, C6JCA007, C6JCA008, D6JCA011, K5JCA041.
  2. Stop using cuvettes from the affected lots and follow your facility's procedures for quarantine.
  3. Contact Accriva Diagnostics for guidance on next steps and any replacement.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette, REF: J103
Sold in
Nationwide
Made for
All ages (0 months and older)

Details

Units affected
28710
Illnesses reported
None reported
Manufactured in
6260 Sequence Dr, San Diego, California, United States
Classification
Class II
Recall number
Z-0317-2027

In the FDA's words

Description
Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette, REF: J103 Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis. UDI-DI: 10711234103012, Lot(Expiration): B6JCA005(02/28/2027), C6JCA008(03/31/2027), D6JCA011(04/30/2027), K5JCA041(10/31/2026), C6JCA007(03/31/2027) Distribution: Worldwide - US Nationwide distribution in the states of MA, MO, AK, WV, NJ, TX, VA, IL, NY and the countries of China, Colombia. Affected quantity: 28710
Hazard
Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis.