Moderate riskFDA DeviceClass II

GE Senographe Pristina VIA recalled for possible extra X-ray dose

GE HealthCare found a nonconformance in certain Senographe Pristina VIA systems. In 3D+2D manual mode, when the anode track and filter combination is configured differently, the collimator blades might show in the image or the collimator might open more than expected, and the wrong collimator filter might be used for the selected anode track. This could deliver a minor additional X-ray dose to the patient and cause obvious image artifacts.

From the FDA notice · Read the official notice ↗ · All GE recalls

What to do

  1. Check whether your facility uses a Senographe Pristina VIA system with one of the GTINs listed in the notice.
  2. Contact GE HealthCare about affected systems. The notice does not specify a fix or correction, so ask the company what steps to take.
  3. The notice does not say how to get a remedy; ask GE HealthCare directly.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Senographe Pristina
Sold in
Nationwide
Made for
Adults (0 months and older)

This recall covers 3 products — see all of them.

Details

Illnesses reported
None reported
Manufactured in
3000 N Grandview Blvd, Waukesha, Wisconsin, United States
Classification
Class II
Recall number
Z-0326-2027

In the FDA's words

Description
Senographe Pristina GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades). GTIN(s): 00840682118460, 00840682142052, 00840682145879, 00195278022745, 00195278276971, 00195278661999, 00195278665379, 00195278914026, 00198953109281, 10198953054311, 10195278901764. Distribution: USA and OUS.
Hazard
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).