FDA-Device2019-11-20Class II

Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.

Canon Medical System, USA, INC.
Hazard

Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.

Sold states
Worldwide distribution - US Nationwide distribution in the states of AL, CT, FL, IL, OR, PA, and TX; and country of Netherlands.
Affected count
9 systems
Manufactured in
2441 Michelle Dr, N/A, Tustin, CA, United States
Products
Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0360-2020

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.