Boston Scientific TCD electrodes recalled for possible contamination
Some TCD radiofrequency electrodes may have small cut marks on the Tyvek side of the sterile pouch. If the sterile barrier is compromised and the devices are used, microbial or other foreign material could enter patient tissue, which may lead to infection, pain, inflammation, allergic reaction, or the need for more medical or surgical treatment. No injuries have been reported.
From the FDA notice · Read the official notice ↗ · All Boston Scientific recalls
What to do
- Do not use affected electrodes. Set aside any pouch with a cut mark or matching lot.
- Affected models are TCD-5, TCD-10, TCD-15, TCD-10-3M and TCD-15-3M, including their multi-pack (-P) versions. Check the lot numbers in the notice.
- The notice doesn't say how to return or replace the devices. Contact Boston Scientific Neuromodulation Corporation for instructions.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- TCD Electrodes (Description/REF(Inner/Outer)): TCD Electrode, Multi-Pack, 3M Cable, 10cm x 28ga (0.36mm)/(TCD-10-3M/TCD-10-3M-P); TCD Electrode, Multi-Pack, 10cm x 28ga (0.36mm)/(TCD-10/TCD-10-P); TCD Electrode, Multi-Pack, 3M Cable, 15cm x 28ga (0.36mm)/(TCD-15-3M/TCD-15-3M-P); TCD Electrode, Multi-Pack, 15cm x 28ga (0.36mm)/(TCD-15/TCD-15-P); TCD Electrode, Multi-Pack, 5cm x 28ga (0.36mm)/(TCD-5/TCD-5-P);
- Sold in
- Nationwide
- Made for
- All ages (under 100 years)
Details
- Units affected
- 13320
- Illnesses reported
- None reported
- Manufactured in
- 25155 Rye Canyon Loop, Valencia, California, United States
- Classification
- Class II
- Recall number
- Z-0368-2027
In the FDA's words
- Description
- TCD Electrodes (Description/REF(Inner/Outer)): TCD Electrode, Multi-Pack, 3M Cable, 10cm x 28ga (0.36mm)/(TCD-10-3M/TCD-10-3M-P); TCD Electrode, Multi-Pack, 10cm x 28ga (0.36mm)/(TCD-10/TCD-10-P); TCD Electrode, Multi-Pack, 3M Cable, 15cm x 28ga (0.36mm)/(TCD-15-3M/TCD-15-3M-P); TCD Electrode, Multi-Pack, 15cm x 28ga (0.36mm)/(TCD-15/TCD-15-P); TCD Electrode, Multi-Pack, 5cm x 28ga (0.36mm)/(TCD-5/TCD-5-P); Radiofrequency probes may have small cut marks on the Tyvek side of the sterile pouch, and if the sterile barrier is compromised and devices are used, this could cause microbial or other foreign material contamination to be introduced into patient tissue, which may lead to infection, pain, inflammation, allergic reaction, toxicity, the need for additional medical and/or surgical intervention. Inner REF/UDI-DI/Lot(Expiration): TCD-10-3M/00813250014396/39749333(7/6/2030), 85009822(7/12/2030); TCD-10/00813250010978/39741131(7/5/2030), 39750861(7/6/2030), 39757187(7/7/2030), 39766608(7/8/2030), 39768774(7/8/2030), 39772208(7/9/2030), 39778427(7/9/2030), 39780344(7/9/2030), 39781367(7/10/2030), 85006486(7/12/2030), 85011925(7/13/2030); TCD-15-3M/00813250014402/39770714(7/8/2030); TCD-15/00813250010985/39741129(7/5/2030), 39746187(7/6/2030), 39780349(7/10/2030), 39783782(7/10/2030); TCD-5/00813250010961/39751579(7/7/2030), 39783599(7/10/2030) Distribution: US Nationwide distribution including in the states of OH, ID, KY, NJ, GA, NY, MO, SC, MN, IA, MI, MA, WI, TX, ME, IN, AR, CO, IL, VT, VA, MS, WA, OR, NV, AZ, UT, FL, WY, CA, OK, NC, TN, MD, DE, LA, HI, CT, WV, AL, NH, ND. Affected quantity: 13320
- Hazard
- Radiofrequency probes may have small cut marks on the Tyvek side of the sterile pouch, and if the sterile barrier is compromised and devices are used, this could cause microbial or other foreign material contamination to be introduced into patient tissue, which may lead to infection, pain, inflammation, allergic reaction, toxicity, the need for additional medical and/or surgical intervention.