Moderate riskFDA DeviceClass II

Medtronic Xomed ventilation tubes recalled for lack of sterilization

Medtronic Xomed is recalling one lot of 5-pack bobbin ventilation tubes (REF 1010201) because they were not sent through the required sterilization process. After implantation, the tubes could cause infection and inflammation, potentially leading to temporary hearing impairment, ear fluid discharge, or the need for antibiotics. The notice reports no injuries.

From the FDA notice · Read the official notice ↗

What to do

  1. Set aside any unused tubes from lot 0231048277 and do not implant them.
  2. Check the 5-pack packaging for UDI 20763000281612 and lot 0231048277.
  3. If a patient received a tube from this lot, contact their care team.
  4. Contact Medtronic Xomed about the affected lot. The notice doesn't say how to get a return or replacement; ask Medtronic Xomed directly.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
BOBBIN VENTILATION TUBE 5PK 1.14 FLPL, REF: 1010201, Lot# 0231048277 (GTIN 20763000281612).
Sold in
Nationwide
Made for
Children (under 18 years)

This recall covers 2 products — see all of them.

Details

Units affected
2000
Illnesses reported
None reported
Manufactured in
6743 Southpoint Dr N, Jacksonville, Florida, United States
Classification
Class II
Recall number
Z-0388-2027

In the FDA's words

Description
BOBBIN VENTILATION TUBE 5PK 1.14 FLPL, REF: 1010201, Lot# 0231048277 (GTIN 20763000281612). Ventilation tubes, that provide a passageway for movement of air between the auditory canal and the middle ear, were not subjected to the required sterilization process, which after implantation could cause infection and subsequent inflammation potentially leading to temporary hearing impairment, otorrhea (fluid discharge) and the need for antibiotics. UDI-DI: 20763000281612. Lots: 0231048277 Distribution: Worldwide - US Nationwide distribution including in the states of NY, AL, VA, MI, AZ, WA, TX, MA, IL, ME, NM, CA, IN, NC, OH, MO, MN, NH, IA, NV, GA, PA, FL, TN, WV, SC, OK, NE, OR, SD, AK, LA, CO, WI, MD, KS, UT, ID, WY, NJ and the countries of Canada, Australia, New Zealand, Hong Kong. Affected quantity: 2000
Hazard
Ventilation tubes, that provide a passageway for movement of air between the auditory canal and the middle ear, were not subjected to the required sterilization process, which after implantation could cause infection and subsequent inflammation potentially leading to temporary hearing impairment, otorrhea (fluid discharge) and the need for antibiotics.

2 products in this recall

Medtronic Xomed ear ventilation tubes recalled for sterilization failure

Medtronic Xomed recalled two ear ventilation tube products, the Sheehy Ventilation Tube 5PK Button 1.27 FLPL and the Bobbin Ventilation Tube 5PK 1.14 FLPL, because the tubes were not subjected to the required sterilization process. After implantation, this could cause infection and inflammation, potentially leading to temporary hearing impairment, ear fluid discharge, and the need for antibiotics. The notice reports no injuries.

The agency lists each product separately. Check every one you might have.

Related topics