TransMedics OCS liver system cassettes recalled for potential leaks
Some OCS Solution Delivery Cassettes were assembled with syringes made with an atypical lubricant. This results in sub-par performance and has the potential to cause leaks. No injuries have been reported.
From the FDA notice · Read the official notice ↗ · All TransMedics recalls
What to do
- Set aside affected cassettes and contact TransMedics and your care team before using them.
- Check lot numbers 4460-4-737189, 4460-4-738008, 4460-4-738009 and 4460-4-739728 on OCS Solution Delivery Cassettes, catalog 1467.
- The notice doesn't say how to get a replacement or return; ask TransMedics, Inc. directly.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- OCS Liver System for organ transport OCS Solution Delivery Cassettes Model/Catalog Number: 1467 Product Description: OCS Solution Delivery Cassettes The recalled product is a standard off the shelf syringe. It is used as part of solution delivery cassette (SDS Cassette) to be used with the perfusion module of OCS Liver System. Each OCS Solution Delivery Cassettes contains three SDS cassettes. The BD syringe is a component of the SDS Cassette PN 100004459
- Sold in
- Nationwide
- Made for
- All ages (under 100 years)
This recall covers 3 products — see all of them.
Details
- Units affected
- 55 OCS Solution Delivery Cassettes
- Illnesses reported
- None reported
- Manufactured in
- 200 Minuteman Rd Ste 302, Andover, Massachusetts, United States
- Classification
- Class II
- Recall number
- Z-0391-2027
In the FDA's words
- Description
- OCS Liver System for organ transport OCS Solution Delivery Cassettes Model/Catalog Number: 1467 Product Description: OCS Solution Delivery Cassettes The recalled product is a standard off the shelf syringe. It is used as part of solution delivery cassette (SDS Cassette) to be used with the perfusion module of OCS Liver System. Each OCS Solution Delivery Cassettes contains three SDS cassettes. The BD syringe is a component of the SDS Cassette PN 100004459 Faulty Solution Delivery System (SDS) Cassettes within the device set were assembled with syringes that were manufactured with an atypical lubricant, which results in sub-par performance and has the potential to cause leaks. Lot Code: Lot: 4460-4-737189; Expiration Date: 2028-10-08; UDI: +B00944605/$$74460-4-737189/14D20281008/Q3L; Lot: 4460-4-738008; Expiration Date: 2028-11-07; UDI: +B00944605/$$74460-4-738008/14D20281107/Q3C; Lot: 4460-4-738009; Expiration Date: 2028-11-07; UDI: +B00944605/$$74460-4-738009/14D20281107/Q3D; Lot: 4460-4-739728; Expiration Date: 2029-01-14; UDI: +B00944605/$$74460-4-739728/14D20290114/Q3K; Distribution: Worldwide - US Nationwide distribution in the state of NE and the country of Italy. Affected quantity: 55 OCS Solution Delivery Cassettes
- Hazard
- Faulty Solution Delivery System (SDS) Cassettes within the device set were assembled with syringes that were manufactured with an atypical lubricant, which results in sub-par performance and has the potential to cause leaks.
3 products in this recall
TransMedics OCS Liver System sets recalled for potential leaks
TransMedics is recalling three OCS Liver System products: Liver Perfusion Sets (model 3200), Liver Solution Infusion Sets (model 3457), and Solution Delivery Cassettes (model 1467). The cassettes were assembled with syringes made with an atypical lubricant, which may cause sub-par performance and has the potential to cause leaks. The recall covers 1,179 perfusion sets, 90 infusion sets and 55 cassettes, distributed nationwide in Nebraska and to Italy, and no injuries have been reported.
The agency lists each product separately. Check every one you might have.
- OCS Liver System for organ transport
OCS Liver Perfusion Set
Model/Catalog Number: 3200
Product Description: The recalled product is a standard off the…Z-0389-2027 · 1179 (Liver Perfusion Sets)
- OCS Liver System for organ transport
OCS Liver Solution Infusion Set
Model/Catalog Number: 3457
Product Description: OCS Liver Solution Infusion Set;…Z-0390-2027 · 90 Liver Solution Infusion Sets
- OCS Liver System for organ transport
OCS Solution Delivery Cassettes
Model/Catalog Number: 1467
Product Description: OCS Solution Delivery Cassettes… (this page)Z-0391-2027 · 55 OCS Solution Delivery Cassettes
More from TransMedics
See all- TransMedics OCS Heart Perfusion Set recalled for incorrect components
ModerateFDANov 6, 2026
- TransMedics OCS Heart Perfusion Set recalled for possible startup error
ModerateFDANov 6, 2026
- TransMedics OCS Liver Perfusion Set recalled for incorrect accessories
ModerateFDANov 6, 2026