Moderate riskFDA DeviceClass II

Helena V8 Serum Protein SPE Kit recalled for possible albumin error

The V8 SPE Diluent in this serum protein kit may lower the albumin result and raise the beta and gamma globulin results. Quality control may fail, so no results may be reported and a repeat sample may be needed. An incorrect result could lead to unnecessary clinical workup or hide a mild true elevation in albumin.

From the FDA notice · Read the official notice ↗

Update since first reported

  • Nov 19, 2026Hazard description updated

What to do

  1. Set aside kits matching Catalog No. 1805, Kit Lot 4-26-1805, expiration 2028-05-31, UDI-DI M52518050.
  2. Check for the affected component: Product No. 552512, Lot 3-26-552512 (V8 SPE Diluent).
  3. The notice doesn't say how to get a replacement or refund; contact Helena Laboratories, Corp. for next steps.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
V8 Serum Protein SPE Kit, Cat. No.: 1805, containing V8 SPE Diluent, Product No.: 552512
Sold in
Nationwide
Made for
All ages (0 months and older)

Details

Units affected
121
Illnesses reported
None reported
Manufactured in
1530 Lindbergh Dr, Beaumont, Texas, United States
Classification
Class II
Recall number
Z-0469-2027

In the FDA's words

Description
V8 Serum Protein SPE Kit, Cat. No.: 1805, containing V8 SPE Diluent, Product No.: 552512 Serum protein kit has a diluent that may produce a decrease in the albumin fraction with a corresponding increase in the beta and gamma globulin fractions. Quality control may fail, with no results reported and repeat sample needed. An incorrect report could lead to unnecessary clinical workup or could mask a mild true elevation in albumin. Lot: 122-08988162, Exp Date: 10/20/2026; 122-08998235, Exp Date: 10/22/2026; 122-08999465, Exp Date: 10/22/2026; 122-09007792, Exp Date: 10/24/2026; 122-08990738, Exp Date: 10/21/2026; 122-09001303, Exp Date: 10/22/2026; 122-09020321, Exp Date: 10/28/2026; 122-09022445, Exp Date: 10/28/2026 Distribution: Worldwide - US Nationwide distribution in the states of MI, CA, TN, MS, MO, TX, IN, FL and the countries of PRY, CAN, ISR, PHL, TTO. Affected quantity: 121
Hazard
Serum protein kit has a diluent that may produce a decrease in the albumin fraction with a corresponding increase in the beta and gamma globulin fractions. Quality control may fail, with no results reported and repeat sample needed. An incorrect report could lead to unnecessary clinical workup or could mask a mild true elevation in albumin.

Related topics