Philips EPIQ ultrasound systems recalled for possible control failure
Philips Ultrasound, LLC is recalling EPIQ ultrasound systems because their TGC slide controls, which adjust gain compensation over a 2D image region, may become unresponsive after multiple start-ups or reboots. Rebooting may not restore the controls. The notice reports no injuries.
From the FDA notice · Read the official notice ↗ · All Philips recalls
What to do
- Clinicians should contact Philips Ultrasound about the technical solution released on March 15, 2024 for affected systems.
- Check whether the system's REF number is on the list, such as 795200 through 795263, and whether it is from lot 124624.
- The notice does not describe a refund or replacement; the technical solution is the remedy it gives.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- EPIQ Ultrasound Systems, REF: 795200, 795201, 795204, 795205, 795206, 795231, 795232, 795234, 795236, 795237, 795239, 795240, 795241, 795242, 795244, 795245, 795247, 795248, 795251, 795253, 795257, 795258, 795263.
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
Details
- Units affected
- 808
- Illnesses reported
- None reported
- Manufactured in
- 22100 Bothell Everett Hwy, Bothell, Washington, United States
- Classification
- Class II
- Recall number
- Z-0473-2027
In the FDA's words
- Description
- EPIQ Ultrasound Systems, REF: 795200, 795201, 795204, 795205, 795206, 795231, 795232, 795234, 795236, 795237, 795239, 795240, 795241, 795242, 795244, 795245, 795247, 795248, 795251, 795253, 795257, 795258, 795263. Diagnostic ultrasound systems have slide controls that adjust brightness range of gain compensation over a unique 2D image region, which may become unresponsive after multiple system start-ups/reboots. Rebooting systems may not recover TGC Slide Controls. The technical solution to correct affected systems was released on 15-Mar-2024. Lot: 122-08990672, Exp Date: 10/20/2026; 122-08995502, Exp Date: 10/21/2026 Distribution: US Nationwide distribution including in the states and territory of NY, CT, FL, IL, AL, CA, VA, OH, MN, AZ, IN, KY, MI, WI, MA, MO, WV, NC, GA, DC, TX, TN, PA, MD, KS, MT, HI, NH, SC, AR, ME, NE, WA, NJ, RI, CO, OK, MS, UT, PR and the countries of India, France, China, Mexico, Australia, New Zealand, Japan, Thailand, Spain, South Korea, Switzerland, Taiwan, Poland, Indonesia, Iran, Singapore, Austria, Italy, Germany, Greece, Portugal, Canada, United Kingdom, Egypt, Iceland, South Africa, Romania, Czech Republic, Costa Rica, Brazil, Lebanon, Ireland, Malaysia, Sweden, Netherlands, Slovakia, Bangladesh. Affected quantity: 808
- Hazard
- Diagnostic ultrasound systems have slide controls that adjust brightness range of gain compensation over a unique 2D image region, which may become unresponsive after multiple system start-ups/reboots. Rebooting systems may not recover TGC Slide Controls. The technical solution to correct affected systems was released on 15-Mar-2024.
More from Philips
See all- Philips EPIQ Elite ultrasound recalled for Doppler display error
ModerateFDANov 17, 2026
- Philips EPIQ ultrasound systems recalled for possible software errors
ModerateFDANov 12, 2026
- Philips Lumify ultrasound system recalled for possible connection loss
SeriousFDAOct 16, 2026
- Philips EPIQ Elite ultrasound system recalled for doppler measurement software error
ModerateFDASep 30, 2026