Beta Bionics iLet pump firmware recalled for stopped insulin delivery
Certain iLet ACE Pump firmware versions have a software defect that can stop insulin delivery after about 7 to 11 months of use. When this happens, the pump becomes unresponsive and, after 30 minutes, shows an "iLet Malfunction (57)" alarm. Restarting the pump temporarily fixes the problem, but the defect can come back, and untreated interruptions can lead to high blood sugar and diabetic ketoacidosis.
From the FDA notice · Read the official notice ↗ · All Beta Bionics recalls
What to do
- Do not stop your insulin pump on your own. Contact your diabetes care team and Beta Bionics before changing your pump or insulin plan.
- If the pump stops and shows the "iLet Malfunction (57)" alarm, restart it as the notice describes, and tell your care team.
- Check the model number BB1001, UDI-DI 00850050080015, and software version: Color Rev E v. 3.3.11 or Monochromatic Rev I v. 1.6.6.
- The notice does not say how to get a fix. Contact Beta Bionics for next steps.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Generic Name: iLet Bionic Pancreas Generic Name: iLet ACE Pump Model/Catalog Number: BB1001 UDI-DI code: 00850050080015 Software Version: (Color Rev E v. 3.3.11; Monochromatic Rev I v. 1.6.6.). Product Description: iLet ACE Pump Component: Firmware for iLet ACE Pump
- Sold in
- Nationwide
- Made for
- Adults (0 months and older)
Details
- Units affected
- Color Rev E v. 3.3.11 = 31,303; Monochromatic Rev I v. 1.6.6 = 3,697
- Illnesses reported
- None reported
- Manufactured in
- 11 Hughes, Irvine, California, United States
- Classification
- Class II
- Recall number
- Z-0531-2027
In the FDA's words
- Description
- Generic Name: iLet Bionic Pancreas Generic Name: iLet ACE Pump Model/Catalog Number: BB1001 UDI-DI code: 00850050080015 Software Version: (Color Rev E v. 3.3.11; Monochromatic Rev I v. 1.6.6.). Product Description: iLet ACE Pump Component: Firmware for iLet ACE Pump Certain iLet Bionic Pancreas firmware versions have a software defect that impacts an internal dose counter of the insulin pump after approximately 7 11 months of use and causes the device to stop delivering insulin. When this occurs, the pump becomes unresponsive and, after 30 minutes, triggers an "iLet Malfunction (57)" alarm that disables the pump until it is manually restarted. Restarting the pump temporarily resolves the issue, but the defect will eventually recur. This problem was an unintended consequence of a prior software update that was originally intended to fix a separate bug. If left untreated, interrupted insulin delivery can lead to high blood sugar (hyperglycemia) and diabetic ketoacidosis (DKA). W037925198361 (double collection) Distribution: US Nationwide distribution. Affected quantity: Color Rev E v. 3.3.11 = 31,303; Monochromatic Rev I v. 1.6.6 = 3,697
- Hazard
- Certain iLet Bionic Pancreas firmware versions have a software defect that impacts an internal dose counter of the insulin pump after approximately 7 11 months of use and causes the device to stop delivering insulin. When this occurs, the pump becomes unresponsive and, after 30 minutes, triggers an "iLet Malfunction (57)" alarm that disables the pump until it is manually restarted. Restarting the pump temporarily resolves the issue, but the defect will eventually recur. This problem was an unintended consequence of a prior software update that was originally intended to fix a separate bug. If left untreated, interrupted insulin delivery can lead to high blood sugar (hyperglycemia) and diabetic ketoacidosis (DKA).