Serious riskFDA DeviceClass I

Abbott FreeStyle Libre 3 app recalled for security and signal loss

Abbott Diabetes Care is recalling the FreeStyle Libre 3 app, which may have a security vulnerability. Users may also experience extended signal loss, where the sensor loses connection with the app and does not reconnect.

From the FDA notice · Read the official notice ↗ · All Abbott recalls

What to do

  1. Do not stop using your glucose monitoring on your own; contact your diabetes care team and Abbott Diabetes Care, Inc. about this app.
  2. Check whether your FreeStyle Libre 3 app is version 3.2.0 through 3.6.8.
  3. Affected products include model 72075-01 (UDI-DI 00357599816001) and model 72077-01 (UDI-DI 00357599817008).
  4. The notice doesn't say how to get a fix or update; ask Abbott Diabetes Care, Inc. for guidance.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
FreeStyle Libre 3 App Software versions: Versions 3.2.0 through 3.6.8 Model/Catalog Number: 72077-01 UDI-DI code: 00357599817008 Model/Catalog Number: 72075-01 UDI-DI code: 00357599816001 The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
Sold in
Nationwide
Made for
All ages (4 years to 120 years)

This recall covers 3 products — see all of them.

Details

Units affected
1,836,506 (US 1,382,849; OUS 453,657) app downloads total
Illnesses reported
None reported
Manufactured in
1360 S Loop Rd, Alameda, California, United States
Classification
Class I
Recall number
Z-0536-2027

In the FDA's words

Description
FreeStyle Libre 3 App Software versions: Versions 3.2.0 through 3.6.8 Model/Catalog Number: 72077-01 UDI-DI code: 00357599817008 Model/Catalog Number: 72075-01 UDI-DI code: 00357599816001 The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated. Due to two issues: 1. Security vulnerability issue. 2. Extended Signal Loss - users may experience extended signal loss, meaning their sensor loses connection with the app and does not reconnect. Box Part No. 4230443; Cartridge Part No. 4220443; UDI H66842204430; Lot: 26GP7665; Distribution: Worldwide distribution - US Nationwide and the countries of Andorra, Austria, Belgium, Switzerland, Germany, Denmark, Spain, Finland, France, United Kingdom, Italy, Lichtenstein, Luxembourg, Netherlands, Norway, San Marino, Sweden, Qatar, Saudi Arabia, Australia, New Zealand, and Canada. Affected quantity: 1,836,506 (US 1,382,849; OUS 453,657) app downloads total
Hazard
Due to two issues: 1. Security vulnerability issue. 2. Extended Signal Loss - users may experience extended signal loss, meaning their sensor loses connection with the app and does not reconnect.

3 products in this recall

FreeStyle Libre, Lingo and Libre apps recalled for signal loss

Abbott is recalling three glucose monitoring apps: the FreeStyle Libre 3 App, the Lingo Biosensor App and the Libre App, with 1,836,506 app downloads affected in the US and other countries. The recall is for a security vulnerability and extended signal loss, where a sensor may lose its connection with the app and not reconnect. The FDA has classified it as Class I, and no injuries have been reported.

The agency lists each product separately. Check every one you might have.