Artis Zee and Zeego, model no. 10272462 AXIOM Artis zee/Zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
This recall has been initiated due to a possible leakage of a hose in the cooling unit of the X-ray tube. The Artis zee system may lose coolant due to a possible leakage in a hose of the cooling unit in the X-ray tube. This will cause the X-ray tube not to be sufficiently cooled down. As a result, the system will display a message Tube hot, have a break . The X-ray will be blocked several minutes later and to prevent any damage to the machine the following message will be displayed NO XRAY, TUBE TOO HOT . This issue is very sporadic and may occur during an ongoing clinical procedure. Use of this product may have a potential impact on planned procedures as they may be delayed as well as in emergent cases, as clinical treatment may need to be terminated and transferred to a functioning system.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0558-2019More Siemens Medical Solutions USA, Inc recalls
- FDA-Device2026-07-08Siemens ARTIS icono ceiling monitor suspension recalled for loose screw risk
- FDA-Device2026-07-08Siemens ARTIS icono floor recalled for potential movement loss at detector height limits
- FDA-Device2026-07-08Siemens ARTIS icono biplane recalled for potential loss of movement after detector adjustment
- FDA-Device2026-07-08Siemens ARTIS icono ceiling model 11328100 recalled for loss of movement capability