- 2025-10-08FDA-DeviceClass IMAGNETOM Vida Fit. Model Number: 11410481.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IMAGNETOM Vida (DE). Model Number: 11060815. 510(k): K170396, K173617, K181433, K183254, K192924, K203443, K213693, K213805, K220425, K231560.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IMAGNETOM Verio Dot Upgrade. Model Number: 10684334.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IMAGNETOM Verio Dot. Model Number: 10684333.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IMAGNETOM Verio. Model Number: 10276755.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IMAGNETOM Spectra as the following: 1. MAGNETOM Spectra. Model Number: 10837643. 2. MAGNETOM Spectra (CN). Model Number: 10655588.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IMAGNETOM Skyra Fit BioMatrix. Model Number: 11516217.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IMAGNETOM Skyra fit. Model Number: 10849580.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IMAGNETOM Skyra (DE). Model Number: 10432915. 510(k) Numbers: K101347, K111242, K123510, K133435, K140253, K141977, K142515, K153343, K173592, K202014, K231560.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IMAGNETOM Prisma. Model Number: 10849582.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IMAGNETOM Lumina (DE). Model Number: 11344916.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IMAGNETOM Connectom.X. Model Number: 11371480.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IMAGNETOM Cima.X (DE). Model Number: 11647158.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IBIOGRAPH One (DE). Model Number: 11689172.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-10-08FDA-DeviceClass IBiograph mMR. Model Number: 10433372.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- 2025-09-17FDA-DeviceClass IIArtis Pheno. Image-Intensified Flouroscopic X-Ray System.
Limited system movements after startup .
- 2025-07-09FDA-DeviceClass IIinterventional fluoroscopic x-ray system
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
- 2025-07-09FDA-DeviceClass IIinterventional fluoroscopic x-ray system
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
- 2025-07-09FDA-DeviceClass IIinterventional fluoroscopic x-ray system
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
- 2025-07-09FDA-DeviceClass IIARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e., patient extremities. This
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
- 2025-06-25FDA-DeviceClass IIUROSKOP Omnia Max. Model Number: 10762473
A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.
- 2025-06-25FDA-DeviceClass IIUROSKOP Omnia. Model Number: 10094910
A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.
- 2025-02-05FDA-DeviceClass IIARTIS One Angiographic X-Ray System
A potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose-rate may be higher than intended by the user.
- 2024-10-02FDA-DeviceClass IIACUSON Maple 1.0 Diagnostic Ultrasound System, REF: 11711750, with software. ACUSON Maple Select 1.0 Diagnostic Ultrasound System, REF: 11711779, with software
If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
- 2024-10-02FDA-DeviceClass IIACUSON Redwood 1.0 and 2.0 Diagnostic Ultrasound Systems, REF: 11503314, with software
If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
- 2024-10-02FDA-DeviceClass IIACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON Juniper Select 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11653093, with software.
If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
- 2024-08-14FDA-DeviceClass IIMultitom Rax -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image Model Number: 10860700
The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
- 2024-08-14FDA-DeviceClass IILUMINOS Lotus Max -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image Model Number: 11574100
The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
- 2024-08-14FDA-DeviceClass IILuminos dRF Max- Intended to visualize anatomical structures by converting an x-ray pattern into a visible image. Model Number: 10762471
The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
- 2024-08-14FDA-DeviceClass IILuminos Agile Max -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image. Model Number: 10762472
The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays