FDA-Device2025-12-03Class IIPROCESSING DEFECT
Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table
Hazard
The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
Intended for
Adults
Illnesses reported
None reported
Sold states
Distribution to US nationwide, Canada and Japan.
Affected count
34 units
Manufactured in
3003 9th Ave Sw, N/A, Watertown, SD, United States
Products
Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0587-2026Want us to check the next one for you?
We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.