Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions: " Pulmonary thromboembolism when anticoagulants are contraindicated " Failure of anticoagulant therapy in thromboembolic disease " Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional t
Due to complaints of increased resistance when advancing the dilator within the introducer sheath.
From the FDA notice · Read the official notice ↗ · All Argon Medical Devices recalls
What to do
The agency's notice has the remedy and who to contact.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions: " Pulmonary thromboembolism when anticoagulants are contraindicated " Failure of anticoagulant therapy in thromboembolic disease " Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced " Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated
- Sold in
- Nationwide
- Made for
- Adults (18 years to 100 years)
Details
- Units affected
- Total devices 1821 (US=1794 and O.U.S.=27) units
- Illnesses reported
- None reported
- Manufactured in
- 1445 Flat Creek Rd, N/A, Athens, TX, United States
- Classification
- Class II
- Recall number
- Z-0589-2026
In the FDA's words
- Description
- Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions: " Pulmonary thromboembolism when anticoagulants are contraindicated " Failure of anticoagulant therapy in thromboembolic disease " Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced " Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated Due to complaints of increased resistance when advancing the dilator within the introducer sheath. This issue could potential result in procedural delays and/or risk of venous intimal injuries if excessive force is applied Model Number: 352506070E UDI-DI code: 00886333217151 Lot Numbers: 11633166 11633125 11629543 11629273 11628836 Distribution: U.S. (nationwide) distribution to states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. O.U.S. (international) to countries of: China, Panama and Trinidad and Tobago Affected quantity: Total devices 1821 (US=1794 and O.U.S.=27) units
- Hazard
- Due to complaints of increased resistance when advancing the dilator within the introducer sheath. This issue could potential result in procedural delays and/or risk of venous intimal injuries if excessive force is applied