FDA-Device2018-02-28Class II

Pentax Video Duodenoscope Model: ED-3490TK (UDI of design being recalled: 04961333232420) These instruments are intended to provide optical visualization of (via a video monitor), and therapeutic access to, the biliary tract via the upper GI tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: esophagus, stomach, duodenum, common bile, hepatic, and cystic ducts.

Pentax of America Inc
Hazard

The duodenoscopes are being recalled in order to replace the forceps elevator mechanism, the O-rings, and the distal end covering to be consistent with the updated design as well as provide an updated periodic inspection as part of the Operation Manual in order to mitigate the potential risk of infection in flexible endoscopy.

Sold states
USA (nationwide) Distribution to the states of including Puerto Rico : AL, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, WA, WI, WV and WY, and the District of Columbia.
Affected count
559
Manufactured in
3 Paragon Dr, N/A, Montvale, NJ, United States
Products
Pentax Video Duodenoscope Model: ED-3490TK (UDI of design being recalled: 04961333232420) These instruments are intended to provide optical visualization of (via a video monitor), and therapeutic access to, the biliary tract via the upper GI tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: esophagus, stomach, duodenum, common bile, hepatic, and cystic ducts.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0643-2018