Pentax of America Inc
4 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Pentax of America Inc and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-08-27FDA-DeviceClass IIPentax Medical Video Processor; Model Number: EPK-i8020c;
During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient and may cause thermal injury to the patient s mucous membrane.
- 2025-03-05FDA-DeviceClass IIPentax Medical Video Processor- Intended to be used with PENTAX Medical endoscopes, monitors and other peripheral devices for endoscopic diagnosis, treatment, and video observation. Model: EPK-i8020c
During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.
- 2025-03-05FDA-DeviceClass IIPENTAX Medical Video Upper GI Scope-Intended to provide optical visualization of (via a video monitor), and therapeutic access to the lower gastrointestinal tract Model: EG29-i20c
During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.
- 2025-03-05FDA-DeviceClass IIPENTAX Medical Video Colonoscope- Intended to provide optical visualization of (via a video monitor), and therapeutic access to the lower gastrointestinal tract Model: EC38-i20cL
During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.