FDA-Device2025-01-22Class II

GE Healthcare Discovery NM/CT 670, Model/Catalog Numbers: 1) 5376204-70-54-2, 2) H2401BC, 3) H2401CE, 4) H2401CG, 5) H3100AG, 6) H3100JW, 7) H3100NC; System, Tomography, Computed, Emission

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Hazard

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Sold states
Worldwide
Affected count
859 units
Manufactured in
4, Hayozma St, Tirat Carmel, Israel
Products
GE Healthcare Discovery NM/CT 670, Model/Catalog Numbers: 1) 5376204-70-54-2, 2) H2401BC, 3) H2401CE, 4) H2401CG, 5) H3100AG, 6) H3100JW, 7) H3100NC; System, Tomography, Computed, Emission

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0891-2025