- 2026-01-28FDA-DeviceClass IIGE Healthcare Omni Legend, sold under the following names: OMNI Legend 16, Omni Legend 21, OMNI Legend 32, Omni Legend PET Gantry 16cm - MEA, Omni Legend PET Gantry 16cm Haifa LM, Omni Legend PET Gantry 16cm Mobile, Omni Legend PET Gantry 21cm - BJ for Global, Omni Legend PET Gantry 21cm - Global Haifa LM, Omni Legend PET Gantry 21cm - Mobile, Omni Legend PET Gantry 21cm WSO LM, Omni Legend PET Gantry 32cm - MEA, Omni Legend PET Gantry 32cm Haifa LM, Omni Legend PET Gantry 32cm Mobile, OMNI
There is a potential intermittent issue on certain Omni Legend systems that can result in a streaking artifact in the PET clinical scan images. This streaking artifact is most easily identified in transaxial slices on the acquisition console (both corrected and non-corrected for attenuation).
- 2026-01-21FDA-DeviceClass IIMAXXUS, System, Tomography, Computed, Emission
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
- 2026-01-21FDA-DeviceClass IISMV BodyTrack, System, Tomography, Computed, Emission
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
- 2026-01-21FDA-DeviceClass IISMV DSI, System, Tomography, Computed, Emission
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
- 2026-01-21FDA-DeviceClass IISMV DSX New Line, System, Tomography, Computed, Emission
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
- 2025-10-29FDA-DeviceClass IIStarcam/Camstar; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29FDA-DeviceClass IISopha Medical Vision (SMV) DSXi/DST-XLi,/ DST-XL/ DSTi; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29FDA-DeviceClass IIOptima NX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29FDA-DeviceClass IIMillenium MPS; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29FDA-DeviceClass IIMillenium MPR; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29FDA-DeviceClass IIMillenium Myosight; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29FDA-DeviceClass IIMillenium MG/MC; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29FDA-DeviceClass IIMaxicam; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29FDA-DeviceClass IIElscint Helix; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29FDA-DeviceClass IIElscint Varicam; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29FDA-DeviceClass IIElscint SPX6; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29FDA-DeviceClass IIElscint SPX4; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29FDA-DeviceClass IIElscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29FDA-DeviceClass IIElscint Cardial; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-07-16FDA-DeviceClass IIVaricam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System
Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
- 2025-07-16FDA-DeviceClass IIInfinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4 Nuclear Medicine system
Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.