FDA-Device2019-03-13Class II

C2 CryoBalloon Controller, REF: FG-1017, with Controller Software v1.18.258 The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus.

PENTAX of America Inc
Hazard

The Controller does not detect overpressure in the balloon during the application of non-dosing puffs of Nitrous Oxide, which can contribute to balloon over pressurization, if the intended vent lumen of the catheter is significantly occluded to prevent relieving balloon pressure due to a kinked catheter condition.If a patient is exposed to higher than physiologic pressures, adverse events such as perforation or mucosal laceration may occur.

Sold states
Worldwide Distribution: US (nationwide) to states of:: CA, DC, FL, LA, MA, MD, MN, NC, NJ, NY, OH, PA, SC, TN, TX, VA, WA;; and countries: Canada, Germany, and Netherlands.
Affected count
45
Manufactured in
303 Convention Way, Ste 1, Redwood City, CA, United States
Products
C2 CryoBalloon Controller, REF: FG-1017, with Controller Software v1.18.258 The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0970-2019