FDA-Device2022-04-27Class II

Artis zee biplane, Model Number 10094141

Siemens Medical Solutions USA, Inc
Hazard

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Sold states
US Nationwide.
Affected count
373 systems
Manufactured in
40 Liberty Blvd, Malvern, PA, United States
Products
Artis zee biplane, Model Number 10094141

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0970-2022

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief