FDA-Device2023-02-01Class II
Software : myNeedle Guide 2D license, myNeedle Guide 2D AWP only Workflow license, or myNeedle Guide 3D license. Used on the following CT (Computed tomography systems): SOMATOM go.Now Model 11061610 NA* SOMATOM go.Now Model 11061618 NA* SOMATOM go.Up Model 11061620 SOMATOM go.Up Model 11061628 NA* SOMATOM go.All Model 11061630 SOMATOM go.All Model 11061638 NA* SOMATOM go.Top Model 11061640 SOMATOM go.Top Model 11061648 NA* SOMATOM X.cite Model 11330001 SOMATOME X.
Siemens Medical Solutions USA, Inc
Hazard
In the syngo.CT VA40 and syngo.CT VA50 myNeedle Guide application, the order of the orientation labels head , center, and feet in the axial thick slice segments are erroneously swapped
Sold states
Nationwide
Affected count
29 units U.S.
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
Software : myNeedle Guide 2D license, myNeedle Guide 2D AWP only Workflow license, or myNeedle Guide 3D license. Used on the following CT (Computed tomography systems): SOMATOM go.Now Model 11061610 NA* SOMATOM go.Now Model 11061618 NA* SOMATOM go.Up Model 11061620 SOMATOM go.Up Model 11061628 NA* SOMATOM go.All Model 11061630 SOMATOM go.All Model 11061638 NA* SOMATOM go.Top Model 11061640 SOMATOM go.Top Model 11061648 NA* SOMATOM X.cite Model 11330001 SOMATOME X.ceed Model 11330002 SOMATOM go.Sim Model 11061660 NA* SOMATOM go.Open Pro Model 11061670 NAEOTOM Alpha Model 11330003
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1003-2023Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief