Beckman Coulter UniCel Dxl 800 analyzer recalled for restart communication issue
Contact Beckman Coulter about UniCel Dxl 800 analyzer
Beckman Coulter UniCel Dxl 800 analyzers (part numbers 973100, A71456, A71457, A25288) may fail to restart after shutdown due to a communication issue. When this happens, the analyzer cannot initialize or process patient samples, QCs, or calibrators, which delays test results.
- If you operate one of these analyzers, contact Beckman Coulter immediately for instructions
- Do not attempt repeated restart cycles; contact support first
- Have an alternative testing method ready in case your analyzer is affected
Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.