Medline surgical procedure kits with Stryker Strykeflow 2 Suction Irrigators recalled for leaking and vapor emission
Medline medical procedure convenience kits containing Stryker's Strykeflow 2 Suction Irrigators have a design flaw that allows irrigation solution to leak into the handpiece and battery pack during use. Some devices have leaked fluid or emitted vaporized saline that may resemble smoke during procedures.
From the FDA notice · Read the official notice ↗ · All Medline recalls
What to do
- Check if your surgical kit matches one of the recalled reference numbers (e.g., CDS860237AF, DYNJ0525321O, DYNJ57640F, DYNJ62594D, or others listed in the recall)
- Stop using any affected kit immediately
- Contact Medline Industries or your hospital supply manager for instructions on return or replacement
- Report any leaks or device malfunctions to Medline or the FDA
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Medline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS, REF CDS860237AF; 2) ROBOTIC PROCEDURE, REF CDS982128Q; 3) CUH THORACIC, REF CDS984338M; 4) THORACIC PROCEDURE, REF CDS984640I; 5) DAVINCI ROBOTIC PACK-LF, REF DYNJ0525321O; 6) LAP CHOLE PACK, REF DYNJ57640F; 7) GYN/PROSTATE ROBOTIC PACK, REF DYNJ62594D; 8) RR-DAVINCI PROSTATE ACCESSORY, REF DYNJ63392D; 9) LAPAROSCOPY PACK, REF DYNJ68187B; 10) LAPAROSCOPY PACK, REF DYNJ82368A; 11) ROBOTIC KIDNEY PACK, REF DYNJ85261; 12) DOWNTOWN ROBOTICS, REF DYNJ86178B; 13) DAVINCI LITHOTOMY PACK, REF DYNJ87610; 14) ROBOTIC WHIPPLE PACK, REF DYNJ88433; 15) PROSTATE DAVINCI, REF DYNJ900127G; 16) GENERAL LAPAROSCOPY, REF DYNJ901847M; 17) LAG TLH, REF DYNJ901890N; 18) LITHOTOMY ROBOTIC CDS, REF DYNJ903948S; 19) LAP, REF DYNJ904576B; 20) ROBOTIC, REF DYNJ905147B; 21) ROBOTICS GYN, REF DYNJ905824M; 22) WH DAVINCI TLH PROCEDURE, REF DYNJ905920B; 23) CHN OR BASIC ROBOTIC, REF DYNJ905982I; 24) KIT ROBOTIC RFD, REF DYNJ906311F; 25) ROBOTIC PROSTATE, REF DYNJ906381G; 26) DAVINCI, REF DYNJ907392C; 27) CHIASSON GASTRIC SLEEVE KIT, REF DYNJ907405C; 28) GYN, REF DYNJ908504B; 29) UROLOGY ROBOTIC, REF DYNJ908588I; 30) THORACIC ROBOTICS, REF DYNJ908777A; 31) GENERAL ROBOT, REF DYNJ908969A; 32) KIT ROBOTIC, REF DYNJ909870B; 33) ROBOTIC THORACOSCOPY, REF DYNJ910924.
- Sold in
- Nationwide
- Made for
- All ages (under 100 years)
This recall covers 6 products — see all of them.
Details
- Units affected
- 4536 units
- Illnesses reported
- None reported
- Manufactured in
- 3 Lakes Dr, Northfield, IL, United States
- Classification
- Class II
- Recall number
- Z-1087-2026
In the FDA's words
- Description
- Medline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS, REF CDS860237AF; 2) ROBOTIC PROCEDURE, REF CDS982128Q; 3) CUH THORACIC, REF CDS984338M; 4) THORACIC PROCEDURE, REF CDS984640I; 5) DAVINCI ROBOTIC PACK-LF, REF DYNJ0525321O; 6) LAP CHOLE PACK, REF DYNJ57640F; 7) GYN/PROSTATE ROBOTIC PACK, REF DYNJ62594D; 8) RR-DAVINCI PROSTATE ACCESSORY, REF DYNJ63392D; 9) LAPAROSCOPY PACK, REF DYNJ68187B; 10) LAPAROSCOPY PACK, REF DYNJ82368A; 11) ROBOTIC KIDNEY PACK, REF DYNJ85261; 12) DOWNTOWN ROBOTICS, REF DYNJ86178B; 13) DAVINCI LITHOTOMY PACK, REF DYNJ87610; 14) ROBOTIC WHIPPLE PACK, REF DYNJ88433; 15) PROSTATE DAVINCI, REF DYNJ900127G; 16) GENERAL LAPAROSCOPY, REF DYNJ901847M; 17) LAG TLH, REF DYNJ901890N; 18) LITHOTOMY ROBOTIC CDS, REF DYNJ903948S; 19) LAP, REF DYNJ904576B; 20) ROBOTIC, REF DYNJ905147B; 21) ROBOTICS GYN, REF DYNJ905824M; 22) WH DAVINCI TLH PROCEDURE, REF DYNJ905920B; 23) CHN OR BASIC ROBOTIC, REF DYNJ905982I; 24) KIT ROBOTIC RFD, REF DYNJ906311F; 25) ROBOTIC PROSTATE, REF DYNJ906381G; 26) DAVINCI, REF DYNJ907392C; 27) CHIASSON GASTRIC SLEEVE KIT, REF DYNJ907405C; 28) GYN, REF DYNJ908504B; 29) UROLOGY ROBOTIC, REF DYNJ908588I; 30) THORACIC ROBOTICS, REF DYNJ908777A; 31) GENERAL ROBOT, REF DYNJ908969A; 32) KIT ROBOTIC, REF DYNJ909870B; 33) ROBOTIC THORACOSCOPY, REF DYNJ910924. Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use. REF CDS860237AF: UDI/DI 10198459232626 (EA) 40198459232627 (CS), Lot Number 25FBK870; REF CDS860237AF: UDI/DI 10198459232626 (EA) 40198459232627 (CS), Lot Number 25EBD110; REF CDS982128Q: UDI/DI 10198459008245 (EA) 40198459008246 (CS), Lot Number 25EBC253; REF CDS982128Q: UDI/DI 10198459008245 (EA) 40198459008246 (CS), Lot Number 25DBG393; REF CDS984338M: UDI/DI 10198459148101 (EA) 40198459148102 (CS), Lot Number 25FBQ303; REF CDS984338M: UDI/DI 10198459148101 (EA) 40198459148102 (CS), Lot Number 25EBD631; REF CDS984338M: UDI/DI 10198459148101 (EA) 40198459148102 (CS), Lot Number 25DBM322; REF CDS984640I: UDI/DI 10198459318955 (EA) 40198459318956 (CS), Lot Number 25DBT807; REF DYNJ0525321O: UDI/DI 10198459236907 (EA) 40198459236908 (CS), Lot Number 25DMG099; REF DYNJ0525321O: UDI/DI 10198459236907 (EA) 40198459236908 (CS), Lot Number 25CMJ006; REF DYNJ57640F: UDI/DI 10198459011115 (EA) 40198459011116 (CS), Lot Number 25FDB347; REF DYNJ62594D: UDI/DI 10198459049767 (EA) 40198459049768 (CS), Lot Number 25EBK539; REF DYNJ62594D: UDI/DI 10198459049767 (EA) 40198459049768 (CS), Lot Number 25DBD145; REF DYNJ63392D: UDI/DI 10198459164507 (EA) 40198459164508 (CS), Lot Number 25BLA967; REF DYNJ68187B: UDI/DI 10198459382086 (EA) 40198459382087 (CS), Lot Number 25FMG440; REF DYNJ82368A: UDI/DI 10198459258077 (EA) 40198459258078 (CS), Lot Number 25EMH020; REF DYNJ82368A: UDI/DI 10198459258077 (EA) 40198459258078 (CS), Lot Number 25CMF917; REF DYNJ85261: UDI/DI 10195327483418 (EA) 40195327483419 (CS), Lot Number 25FBK331; REF DYNJ85261: UDI/DI 10195327483418 (EA) 40195327483419 (CS), Lot Number 25EBR783; REF DYNJ86178B: UDI/DI 10198459104800 (EA) 40198459104801 (CS), Lot Number 25EBP320; REF DYNJ86178B: UDI/DI 10198459104800 (EA) 40198459104801 (CS), Lot Number 25EBC981; REF DYNJ86178B: UDI/DI 10198459104800 (EA) 40198459104801 (CS), Lot Number 25DBA834; REF DYNJ86178B: UDI/DI 10198459104800 (EA) 40198459104801 (CS), Lot Number 25CBM884; REF DYNJ87610: UDI/DI 10198459047817 (EA) 40198459047818 (CS), Lot Number 25FBK123; REF DYNJ87610: UDI/DI 10198459047817 (EA) 40198459047818 (CS), Lot Number 25EBF929; REF DYNJ87610: UDI/DI 10198459047817 (EA) 40198459047818 (CS), Lot Number 25CBT135; REF DYNJ88433: UDI/DI 10198459157936 (EA) 40198459157937 (CS), Lot Number 25FMC215; REF DYNJ88433: UDI/DI 10198459157936 (EA) 40198459157937 (CS), Lot Number 25EMF199; REF DYNJ88433: UDI/DI 10198459157936 (EA) 40198459157937 (CS), Lot Number 25EMC621; REF DYNJ900127G: UDI/DI 10193489851588 (EA) 40193489851589 (CS), Lot Number 25FBO994; REF DYNJ900127G: UDI/DI 10193489851588 (EA) 40193489851589 (CS), Lot Number 25EBE439; REF DYNJ901847M: UDI/DI 10198459037863 (EA) 40198459037864 (CS), Lot Number 25BBM364; REF DYNJ901890N: UDI/DI 10198459300592 (EA) 40198459300593 (CS), Lot Number 25FBJ408; REF DYNJ901890N: UDI/DI 10198459300592 (EA) 40198459300593 (CS), Lot Number 25EBJ877; REF DYNJ903948S: UDI/DI 10198459332883 (EA) 40198459332884 (CS), Lot Number 25EBO236; REF DYNJ903948S: UDI/DI 10198459332883 (EA) 40198459332884 (CS), Lot Number 25EBC133; REF DYNJ904576B: UDI/DI 10888277776760 (EA) 40888277776761 (CS), Lot Number 25DMH506; REF DYNJ904576B: UDI/DI 1088
- Hazard
- Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
6 products in this recall
Medline surgical kits recalled for Stryker irrigator leak risk
Medline is recalling six groups of surgical convenience kits, about 59,000 units in total, that include Stryker Strykeflow 2 Suction Irrigators. A 2024 design change lets irrigation solution reach the handpiece and battery pack, which can cause the device to leak or give off vaporized saline that may look like smoke. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359Z-1085-2026 · 96 units
- Medline medical procedure convenience kits labeled as: 1) LAPAROSCOPY CDS-LF, REF CDS860235L; 2) GENERAL LAPAROSCOPIC, REF CDS930223M; 3) LAP CHOLE CD…Z-1086-2026 · 41432 units
- Medline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS, REF CDS860237AF; 2) ROBOTIC PROCEDURE, REF CDS982128Q; 3) CUH THOR… (this page)Z-1087-2026 · 4536 units
- Medline medical procedure convenience kits labeled as: 1) ROBOTIC, REF CDS980523M; 2) ROBOTIC GYN, REF CDS980647K; 3) DAVINCI PROCEDURE, REF CDS980864…Z-1088-2026 · 12584 units
- Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE, REF CDS980754T; 2) GYN LAP HYSTERECTOMY CDS, REF CDS982662Q; 3) DAVINCI HYST…Z-1089-2026 · 1457 units
- Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116GZ-1090-2026 · 12 units
More from Medline
See all- Medline surgical procedure kits recalled for reprocessed components
ModerateFDANov 20, 2026
- Medline Bubble CPAP kit recalled for possible heat damage risk
ModerateFDANov 20, 2026
- Medline catheter kits recalled for possible lubricant contamination
ModerateFDANov 10, 2026