NIPRO Conductivity Standard Solution recalled for inaccurate calibration values
What this means
NIPRO Technical Services' Conductivity Standard Solutions (models 10001, 10011, 10021, 10032, 10040, 10040c, 10041) have a defect that causes incorrect conductivity readings. These solutions are used to calibrate meters that measure dialysate (the fluid used in dialysis treatment). If calibration is inaccurate, dialysis machines may report wrong conductivity values, which could affect treatment safety.
- Stop using affected NIPRO Conductivity Standard Solution immediately
- Check your inventory for model numbers 10001, 10011, 10021, 10032, 10040, 10040c, or 10041 with lot number 251030
- Contact NIPRO Technical Services for a replacement or refund
- Recalibrate your conductivity meters with a replacement solution from an approved supplier
The agency's original wording
due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.