FDA-Device2026-02-04Class IIPROCESSING DEFECT

NIPRO Conductivity Standard Solution recalled for inaccurate calibration values

OTHERNationwide distribution

Stop using NIPRO Conductivity Standard Solution

NIPRO Technical Services' Conductivity Standard Solutions (models 10001, 10011, 10021, 10032, 10040, 10040c, 10041) have a defect that causes incorrect conductivity readings. These solutions are used to calibrate meters that measure dialysate (the fluid used in dialysis treatment). If calibration is inaccurate, dialysis machines may report wrong conductivity values, which could affect treatment safety.

  • Stop using affected NIPRO Conductivity Standard Solution immediately
  • Check your inventory for model numbers 10001, 10011, 10021, 10032, 10040, 10040c, or 10041 with lot number 251030
  • Contact NIPRO Technical Services for a replacement or refund
  • Recalibrate your conductivity meters with a replacement solution from an approved supplier
Hazard

due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.

Sold states
U.S. Nationwide distribution in the states of AL, AR, AZ, CA, DE, FL, GA. IL, IN, KS, MI, MO, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN TX, UT, VA, and WA.
Affected count
499 units
Manufactured in
3110 N Oakland Ste 101, N/A, Mesa, AZ, United States
Products
Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity Standard Solutions are a secondary standard solution used for the calibration of conductivity cells together with conductivity meters

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1185-2026