Moderate riskFDA DeviceClass II

Encore Medical EMPOWR Acetabular System liners recalled for packaging mismatch

Encore Medical discovered that some packages of EMPOWR Acetabular System hip implant liners contained the wrong device inside—for example, a box labeled as one size contained a different size. This packaging error could force surgeons to delay surgery or perform additional procedures to swap out the correct implant.

From the FDA notice · Read the official notice ↗ · All Encore Medical recalls

What to do

  1. If you work at a hospital or surgical facility, locate any EMPOWR Acetabular System liners (ref 942-01-36H, 942-01-36J, 942-01-40G, or 942-01-32C) with expiration 03-15-2026
  2. Open and verify the contents match the package label before surgery
  3. Contact Encore Medical or your distributor if contents do not match the label
  4. Do not use a mismatched device; return it to Encore Medical

Is yours affected?

Compare your label or packaging with the recalled product.

Product
EMPOWR Acetabular System: Liner, 10 degree Hooded, HXe+, 36H, REF: 942-01-36H; Liner, 10 degree Hooded, HXe+, 36J, REF: 942-01-36J; Liner, 10 degree Hooded, HXe+, 40G, REF: 942-01-40G; Liner, 10 degree Hooded, HXe+, 32C, REF: 942-01-32C
Sold in
US: TX, IA, MI, MS, MA, RI, NJ, PA, NY, MO, OK, PR, AL, CA
Made for
Adults (0 months and older)

This recall covers 2 products — see all of them.

Details

Units affected
49
Illnesses reported
None reported
Manufactured in
9800 Metric Blvd, N/A, Austin, TX, United States
Classification
Class II
Recall number
Z-1234-2026

In the FDA's words

Description
EMPOWR Acetabular System: Liner, 10 degree Hooded, HXe+, 36H, REF: 942-01-36H; Liner, 10 degree Hooded, HXe+, 36J, REF: 942-01-36J; Liner, 10 degree Hooded, HXe+, 40G, REF: 942-01-40G; Liner, 10 degree Hooded, HXe+, 32C, REF: 942-01-32C 942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery. REF/UDI-DI/Lot(Expiration): 942-01-36H/00190446309789/624Z1032(03-15-2026), 942-01-36J/00190446309802/626Z1021(03-15-2026), 942-01-40G/00190446309826/628Z1034(03-15-2026), 942-01-32C/00190446309697/615Z1026(03-15-2026) Distribution: US: TX, IA, MI, MS, MA, RI, NJ, PA, NY, MO, OK, PR, AL, CA Affected quantity: 49
Hazard
942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery.

2 products in this recall

Encore Medical EMPOWR implants recalled for packaging discrepancy

Encore Medical is recalling two EMPOWR surgical device products, 68 units in total, because some packages may contain the wrong acetabular system or may have a packaging discrepancy. One acetabular system package was found to contain a different device, and other acetabular system and knee insert devices may be affected. The problem could delay surgery while devices are exchanged or lead to a revision surgery. No injuries have been reported.

The agency lists each product separately. Check every one you might have.