Encore Medical EMPOWR Acetabular System liners recalled for packaging mismatch
Encore Medical discovered that some packages of EMPOWR Acetabular System hip implant liners contained the wrong device inside—for example, a box labeled as one size contained a different size. This packaging error could force surgeons to delay surgery or perform additional procedures to swap out the correct implant.
From the FDA notice · Read the official notice ↗ · All Encore Medical recalls
What to do
- If you work at a hospital or surgical facility, locate any EMPOWR Acetabular System liners (ref 942-01-36H, 942-01-36J, 942-01-40G, or 942-01-32C) with expiration 03-15-2026
- Open and verify the contents match the package label before surgery
- Contact Encore Medical or your distributor if contents do not match the label
- Do not use a mismatched device; return it to Encore Medical
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- EMPOWR Acetabular System: Liner, 10 degree Hooded, HXe+, 36H, REF: 942-01-36H; Liner, 10 degree Hooded, HXe+, 36J, REF: 942-01-36J; Liner, 10 degree Hooded, HXe+, 40G, REF: 942-01-40G; Liner, 10 degree Hooded, HXe+, 32C, REF: 942-01-32C
- Sold in
- US: TX, IA, MI, MS, MA, RI, NJ, PA, NY, MO, OK, PR, AL, CA
- Made for
- Adults (0 months and older)
This recall covers 2 products — see all of them.
Details
- Units affected
- 49
- Illnesses reported
- None reported
- Manufactured in
- 9800 Metric Blvd, N/A, Austin, TX, United States
- Classification
- Class II
- Recall number
- Z-1234-2026
In the FDA's words
- Description
- EMPOWR Acetabular System: Liner, 10 degree Hooded, HXe+, 36H, REF: 942-01-36H; Liner, 10 degree Hooded, HXe+, 36J, REF: 942-01-36J; Liner, 10 degree Hooded, HXe+, 40G, REF: 942-01-40G; Liner, 10 degree Hooded, HXe+, 32C, REF: 942-01-32C 942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery. REF/UDI-DI/Lot(Expiration): 942-01-36H/00190446309789/624Z1032(03-15-2026), 942-01-36J/00190446309802/626Z1021(03-15-2026), 942-01-40G/00190446309826/628Z1034(03-15-2026), 942-01-32C/00190446309697/615Z1026(03-15-2026) Distribution: US: TX, IA, MI, MS, MA, RI, NJ, PA, NY, MO, OK, PR, AL, CA Affected quantity: 49
- Hazard
- 942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery.
2 products in this recall
Encore Medical EMPOWR implants recalled for packaging discrepancy
Encore Medical is recalling two EMPOWR surgical device products, 68 units in total, because some packages may contain the wrong acetabular system or may have a packaging discrepancy. One acetabular system package was found to contain a different device, and other acetabular system and knee insert devices may be affected. The problem could delay surgery while devices are exchanged or lead to a revision surgery. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: 342-14-705Z-1233-2026 · 19
- EMPOWR Acetabular System: Liner, 10 degree Hooded, HXe+, 36H, REF: 942-01-36H; Liner, 10 degree Hooded, HXe+, 36J, REF: 942-01-36J; Liner, 10 degree Hooded,… (this page)Z-1234-2026 · 49