FDA-Device2023-03-22Class II
Access Immunoassay Systems (1) Access 2 Immunoassay Analyzer (81600N); (2) Access 2 Immunoassay Analyzer, Refurbished (386220); (3) Access 2 Immunoassay Analyzer, Refurbished (A65531) ; and (4) Access 2 Section, DxC 600i Packaged(A25640).
Hazard
An increased incidence of damage to the nut band from a new vendor (Boamax) has led to an increased volume of complaints from consignees regarding wash arm motion failures, system check failures, and/or assay performance leading to a remote probability of delayed results or reporting erroneous patient results. The subject nut bands were found to not meet the firm's specifications for hardness and had the incorrect material grain direction. Affected instruments were manufactured from December 28, 2017 to October 19, 2022 or had their nut bands replaced during the same period.
Sold states
Worldwide and Nationwide Distribution. There was government/military distribution. Foreign distribution was made to Algeria, Andorra, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Bonaire, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Ethiopia, France, Georgia, Germany, Ghana, Gibraltar, Greece, Guyana, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Kyrgyzstan, Lebanon, Libya, Lithuania, Macao, Malawi, Malaysia, Maldives, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, U.A.E, Uganda, Ukraine, United Kingdom, Uruguay, Vietnam, Yemen, Zambia, and Zimbabwe.
Affected count
4,137 devices (990 U.S. and 3,147 OUS)
Manufactured in
1000 Lake Hazeltine Dr, N/A, Chaska, MN, United States
Products
Access Immunoassay Systems (1) Access 2 Immunoassay Analyzer (81600N); (2) Access 2 Immunoassay Analyzer, Refurbished (386220); (3) Access 2 Immunoassay Analyzer, Refurbished (A65531) ; and (4) Access 2 Section, DxC 600i Packaged(A25640).
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1257-2023More Beckman Coulter, Inc. recalls
- FDA-Device2026-04-15Beckman Coulter MicroScan Neg MIC 3J antibiotic susceptibility panels recalled for manufacturing defects
- FDA-Device2026-04-15MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
- FDA-Device2026-04-01Beckman Coulter MicroScan Neg MIC 56 panels recalled for potential well contamination
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief