Serious riskFDA DeviceClass II

Intuitive Surgical da Vinci 5 Surgeon Console display recalled for firmware defect causing vision loss

Certain da Vinci 5 Surgeon Console Viewer Displays have outdated firmware that may cause loss of the 3D surgical display or both displays entirely during surgery. When this happens, the surgical instruments remain active even though the surgeon cannot see the surgical field, which could result in unintended tissue contact, bleeding, or injury.

From the FDA notice · Read the official notice ↗ · All Intuitive Surgical recalls

What to do

  1. If your facility operates a da Vinci 5 with the affected serial numbers (SQ0326–SQ0470 range), contact Intuitive Surgical immediately
  2. Do not use the affected console for surgery until the firmware is updated
  3. Have Intuitive Surgical perform a firmware update on your display assembly

Is yours affected?

Compare your label or packaging with the recalled product.

Product
da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45
Sold in
Nationwide
Made for
All ages (0 months and older)

Details

Units affected
47
Illnesses reported
None reported
Manufactured in
1266 Kifer Rd, N/A, Sunnyvale, CA, United States
Classification
Class II
Recall number
Z-1355-2026

In the FDA's words

Description
da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45 Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality. UDI: 00886874119747, Serial: SQ0326, SQ0336, SQ0341, SQ0345, SQ0346, SQ0356, SQ0367, SQ0368, SQ0375, SQ0377, SQ0379, SQ0380, SQ0386, SQ0387, SQ0389, SQ0394, SQ0395, SQ0398, SQ0401, SQ0404, SQ0412, SQ0413, SQ0415, SQ0419, SQ0420, SQ0422, SQ0425, SQ0427, SQ0428, SQ0432, SQ0433, SQ0436, SQ0443, SQ0447, SQ0448, SQ0451, SQ0454, SQ0457, SQ0458, SQ0462, SQ0464, SQ0469, SQ0470, SQ0662, SQ0727, SQ0740, SQ0761 Distribution: US Nationwide distribution in the states of KS, AZ, MA, FL, MI, NC, IL, DC, TX, KY, MS, CA, MN, NV, NY, WI, ID, GA, CT, SC, UT, VA. Affected quantity: 47
Hazard
Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.

Related topics

OtherSoftwareMedical devicesSold nationwideMore from FDA Device