FDA-Device2022-08-03Class I
Intraosseous needle and driver. Catalog/Device Name: D015151NK/BD Needle Kit for Powered Driver 15mm x 15Ga, D015151MK/ BD Manual Driver Needle Kit 15mm x 15Ga, D015251NK/ BD Needle Kit for Powered Driver 25mm x 15Ga, D015251MK/ BD Manual Driver Needle Kit 25mm x 15Ga, D015351NK/ BD Needle Kit for Powered Driver 35mm x 15Ga, D015351MK/ BD Manual Driver Needle Kit 35mm x 15Ga, D015451NK/ BD Needle Kit for Powered Driver 45mm x 15Ga, D015451MK/ BD Manual Driver Needle Kit 45mm x 15Ga,
Hazard
BD Intraosseous Needle Kits may exhibit: 1) Increased force required to remove the stylet from intraosseous needle during placement may cause removal of the entire needle assembly and loss of intraosseous access; 2) Stylet safety mechanism may not engage as the stylet is removed; 3) Metal discs in the powered driver may stick rendering the driver unusable.
Sold states
Distribution US nationwide and Canada.
Affected count
40,734
Manufactured in
605 N 5600 W, Salt Lake City, UT, United States
Products
Intraosseous needle and driver. Catalog/Device Name: D015151NK/BD Needle Kit for Powered Driver 15mm x 15Ga, D015151MK/ BD Manual Driver Needle Kit 15mm x 15Ga, D015251NK/ BD Needle Kit for Powered Driver 25mm x 15Ga, D015251MK/ BD Manual Driver Needle Kit 25mm x 15Ga, D015351NK/ BD Needle Kit for Powered Driver 35mm x 15Ga, D015351MK/ BD Manual Driver Needle Kit 35mm x 15Ga, D015451NK/ BD Needle Kit for Powered Driver 45mm x 15Ga, D015451MK/ BD Manual Driver Needle Kit 45mm x 15Ga, D015551NK/ BD Needle Kit for Powered Driver 55mm x 15Ga, D015551MK/ BD Manual Driver Needle Kit 55mm x 15G, D001001/ BD Intraosseous Powered Driver (drill)
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1397-2022More Bard Access Systems, Inc. recalls
- FDA-Device2026-07-08CK000495A Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741121609 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 The Poly Per-Q-Cath Midline catheters are indicated for short or long term peripheral access to the peripheral venous system for selected intravenous therapies and blood sarnpling.
- FDA-Device2026-07-085855240 Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107177 5855300 Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107184 5865120 Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107191 5865200 Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI_DI Code: 0080
- FDA-Device2026-07-08CK000727 PowerMidline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741138478 CK000729 PowerMidline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741138713 P4153108D PowerMidline Catheter 3 Fr Single-lumen UDI-DI Code: 00801741108747 P4154108 PowerMidline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108754 P4154108D PowerMidline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108761 The PowerMidline" Catheter is intended for short term peripheral access for selected intrav
- FDA-Device2026-07-08CK000880 Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741188770 S4153108BDP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154003 S4153108BP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741153990 S4153108D Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741153952 S4153108DCP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741233258 S4153108DGP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 008
- FDA-Device2026-07-085625240 DuoGlide Short-Term Straight Dialysis Catheter 13 Fr Dual-Lumen UDI-DI Code: 00801741042119 The intended use of the Short-Term Dialysis Catheter is attaining temporary/short-term (less than 30 days) vascular access for hemodialysis, hemoperfusion, or apheresis therapy via the jugular, subclavian, or femoral vein.
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief