Brand
Bard Access Systems, Inc.
1 federal recall on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Bard Access Systems, Inc. and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-04-23FDA-DeviceClass IBD 4 Fr Single-Lumen PowerPICC Catheters, REF: 3174155, CK000540, 3174118, CK000483, 3174335, 3174355, CK000196, 9174118, CK000955, 3174135, CK000617, CK000839, 3194355, 6174118, 3194135, 3194335, 2174108, 9194118, 3194118, 1174118, CK000285C, CK000335A, CK000796, 1194108D, CK000757, CK001042, 9174108, 1174108D1, CK000631B, 3194155, 3174108, CK000567, 1194108D4, 1174108D2, CK000677, CK000864, CK000937A, CK000930, CK001027, 1174108D, CK000867, CK000900, CK000315A, CK000589A, CK000854B, CK000897,
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Get notified about new Bard Access Systems, Inc. recalls
Free weekly digest. We email when a new federal recall affects Bard Access Systems, Inc. or anything else in your household.
Get the Sunday BriefExplore other recall topics