FDA-Device2023-05-03Class II

FLEX. GRASP. FORCEPS 6.6FR WL 550MM, Product Number 8736.685

Richard Wolf GmbH
Hazard

Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.

Sold states
Worldwide distribution - US Nationwide and the countries of United Arab Emirates (AE), Argentina (AR), Austria (AT), Belgium (BE), BO, Brazil (BR), Canada (CA), Chile (CL), Czech Republic (CZ), Germany (DE), Denmark (DK), Ecuador (EC), Spain (ES), Finland (FI), France (FR), Greece (GR), Hong Kong (HK), Croatia (HR), Ireland (IE), India (IN), Italy (IT), Jordan (JO), Japan (JP), South Korea (KR), Kuwait (KW), Kazakhstan (KZ), LB, Lithuania (LT), Marzocco (MA), Mexico (MX), Netherlands (NL), Norway (NO), New Zealand (NZ), Oman (OM), Pakistan (PK), Poland (PL), Palestine (PS), Portugal (PT), Qatar (QA), Romania (RO), Serbia (RS), Russian Federation (RU), Saudi Arabia (SA), Slovenia (SI), Slovakia (SK), Thailand (TH), Turkey (TR).
Affected count
280 units
Manufactured in
Pforzheimer Str. 32, N/A, Knittlingen, N/A, Germany
Products
FLEX. GRASP. FORCEPS 6.6FR WL 550MM, Product Number 8736.685

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1425-2023